Semiquantitative reverse transcription of polymerase chain reaction analysis for detection of bcr/abl rearrangement using RNA extracts from bone marrow aspirates compared with glass slide smears after 0, 2, and 4 d of storage.

Schoch R, Pitako J, Schafhausen P, Jenisch S, Haferlach T, Kneba M, Suttorp M

Publication: Br J Haematol (2001) • Vol 115 • Page 583PMID: 11736939View on PubMed

Purpose

To validate a nested RT-PCR approach that allows use of aired-dried bone marrow smears for the detection of chromosomal aberrations, to assess the effects of room temperature specimen storage, and to compare results with those attained using bone marrow aspirate specimens.

Conclusion

Results of bcr/abl transcript levels after storage at room temperature for 96 h were comparable among bone marrow aspirates and air-dried smear samples. The authors conclude that diagnosis of chronic myeloid leukemia is not significantly impaired by sample storage at room temperature for up to 96 h.

Studies (2)

Purpose

To assess the sensitivity of the nested RT-PCR assay.

Summary of Findings

The detection threshold of the two-tiered nested RT-PCR assay for the bcr/abl fusion transcript was one malignant in 1 x 10^5 cells.
  • Biospecimens

    CellBone Marrow
  • Preservation Types

    None (Fresh)
  • Diagnoses

    NeoplasticLeukemia
  • Platforms

    RNART-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCell number
    1x10^3 to 1x10^7 cells
    RT-PCR SpecificTargeted nucleic acid
    bcr/abl
    beta-actin