Ultrasound-accelerated formalin fixation of tissue improves morphology, antigen and mRNA preservation

Chu Wei-Sing, Furusato Bungo, Wong Kondi, Sesterhenn Isabell A, Mustofi Fathollah K, Wei Min Q, Zhu Zhenqing, Abbondanzo S, Liang Qi

Publication: Mod Pathol (2005) • Vol 18 • Page 850PMID: 15605077View on PubMed

Purpose

Comparative analysis of the preservation of RNA integrity, tissue morphology, and protein antigenicity among tissues preserved by ultrasound-accelerated formalin fixation (USAFF) and conventional formalin fixation.

Conclusion

Tissue fixation in the presence of ultrasound substantially reduced fixation duration (from 16 h to 5 min) without compromising tissue morphology. Further, superior results were observed for immunohistochemistry, in situ hybridization, Western blot, and RT-PCR analyses when compared to conventional fixation methods.

Studies (5)

Purpose

To assess if tissue morphology is adequately preserved by USAFF (5-30 min) compared to conventional formalin fixation (16-24 h) using several tissues (brain, tonsil, lymph node, lung, prostate, soft tissue) obtained by biopsy or post mortem.

Summary of Findings

Brain and lymphoid tissue fixed for 15-30 min by USAFF yielded tissue morphology comparable to conventional formalin fixation. Further, USAFF for 15 min resulted in enhanced eosin staining compared to other fixation durations and methods.

  • Biospecimens

    TissueBrain
    TissueTonsil
    TissueLymph Node
    TissueLung
    TissueProstate
    TissueOther
  • Preservation Types

    Formalin
  • Diagnoses

    Not specified
    Autopsy
  • Platforms

    MorphologyH-and-E microscopy
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationMethod of fixative delivery
    Immersion
    Ultrasound-accelerated
    Biospecimen PreservationTime in fixative
    15 min
    16 h
    30 min