This paper compared the yield and integrity of DNA and RNA obtained from case-matched unstained formalin-fixed, paraffin-embedded (FFPE) sections stored as sections at room temperature, 4°C, −30°C, and −80°C for 0 days, 1 week, 1 month, 3 months, and 1 year; additionally, unmatched FFPE blocks stored for <1 year and those stored for > 1 year were compared for storage-related effects. DNA integrity was assessed using DNA integrity numbers (DIN), fragment length and ΔΔCq generated using the NGS FFPE QC Kit. RNA integrity was assessed using RNA integrity numbers (RIN) and percentage of fragments >200 nt (DV200).
Preanalytical Determinants of DNA and RNA Quality in FFPE Tissues: Practical Recommendations for Molecular Testing.
Kunita A, Ushiku T, Nagumo S, Shinozaki-Ushiku A
Purpose
Conclusion
DIN and fragment length were very strongly correlated with each other (r=0.9870, P<0.0001) and were strongly and inversely correlated with ΔΔCq, which was generated using the NGS FFPE QC Kit (r=-0.7417, P<0.001 and r=-0.7617, P<0.0001, respectively); however, RIN and DV200 were not significantly correlated. While DNA and RNA yields and RIN were largely unaffected by the storage of unstained sections at room temperature for up to 1 year, DIN was significantly reduced after ≥ 1 week at room temperature (P<0.05), fragment length and DV200 were significantly reduced after 3 months at room temperature (P<0.001 and P<0.01, respectively), and ΔΔCq increased (signifying an increase in degradation) after 12 months at room temperature (P<0.05, both). DNA yields were higher when sections were stored at −30°C versus room temperature for ≥ 3 months (P<0.01). In contrast, RNA yields were lower in sections stored for 1 year at −30°C or -80°C than those stored at room temperature (P<0.05, all). Storage of sections at ≤4°C rather than room temperature prevented declines in DIN, and fragment length, such that both were higher after ≥ 1 month when stored at ≤4°C than at room temperature (significance dependent on storage temperature and duration). Similarly, DV200 was higher after ≥ 3 months when sections were stored at ≤4°C than at room temperature (P<0.001, all), but RIN was unaffected by storage temperature. Unmatched FFPE blocks that were stored for <1 year had higher DNA and RNA yields than those stored for ≥1 year, and blocks stored for ≤1 year rather than >1 years had more rapid declines in fragment length (P=0.0179), RIN (P=0.0241) and DV200 (P=0.0568) with section storage based on a mixed‐effects model. Importantly, the slide storage associated decline in DV200 was also larger in high-quality (DV200 >60%) specimens, such that DV200 was comparable in high-quality and low-quality specimens after 1 year of slide storage.
Studies (1)
Purpose
This study compared the yield and integrity of DNA and RNA obtained from case-matched unstained formalin-fixed, paraffin-embedded (FFPE) sections stored at room temperature, 4°C, −30°C, and −80°C for up to 1 year (0 days, 1 week, 1 month, 3 months, and 1 year); additionally, unmatched FFPE blocks stored for <1 year and those stored for > 1 year were compared. Twenty FFPE specimens (3 liver, 11 colon, 3 pancreas, 2 ovary and 1 uterus specimen) were obtained from the hospital archive (further details were not provided). Ten of the colon specimens were stored for <1 year, but all of the remaining specimens had been stored for 1-5 years. Fresh sections of each block were stored in the dark at room temperature (20°C–22°C), 4°C, −30°C, and −80°C for 0 days, 1 week, 1 month, 3 months, and 1 year. Sections were brought to room temperature before DNA extraction using the QIAamp DNA FFPE Advanced Kit and RNA extraction with the RNeasy FFPE Kit. DNA was quantified using the Qubit dsDNA assay. DNA integrity was evaluated using an Agilent 4200 TapeStation (DIN and fragment length) and by the real-time PCR amplification of 42 bp and 123 bp amplicons using the Agilent NGS FFPE QC Kit. RNA was extracted using the RNeasy FFPE Kit and was quantified using a NanoDrop 2000 with the Qubit RNA Assay. RNA integrity was evaluated using an Agilent 2100 Bioanalyzer.
Summary of Findings
DIN and fragment length were very strongly correlated with each other (r=0.9870, P<0.0001) and were strongly and inversely correlated with ΔΔCq, which was generated using the NGS FFPE QC Kit (r=-0.7417, P<0.001 and r=-0.7617, P<0.0001, respectively); however, RIN and DV200 were not significantly correlated. While DNA and RNA yields and RIN were largely unaffected by the storage of unstained sections at room temperature for up to 1 year, DIN was significantly reduced after ≥ 1 week at room temperature (P<0.05), fragment length and DV200 were significantly reduced after 3 months at room temperature (P<0.001 and P<0.01, respectively), and ΔΔCq increased (signifying an increase in degradation) after 12 months at room temperature (P<0.05, both). DNA yields were higher when sections were stored at −30°C versus room temperature for ≥ 3 months (P<0.01). In contrast, RNA yields were lower in sections stored for 1 year at −30°C or -80°C than those stored at room temperature (P<0.05, all). Storage of sections at ≤4°C rather than room temperature prevented declines in DIN, and fragment length, such that both were higher after ≥ 1 month when stored at ≤4°C than at room temperature (significance dependent on storage temperature and duration). Similarly, DV200 was higher after ≥ 3 months when sections were stored at ≤4°C than at room temperature (P<0.001, all), but RIN was unaffected by storage temperature. Unmatched FFPE blocks that were stored for <1 year had higher DNA and RNA yields than those stored for ≥1 year, and blocks stored for ≤1 year rather than >1 years had more rapid declines in fragment length (P=0.0179), RIN (P=0.0241) and DV200 (P=0.0568) with section storage based on a mixed‐effects model. Importantly, the slide storage associated decline in DV200 was also larger in high-quality (DV200 >60%) specimens, such that DV200 was comparable in high-quality and low-quality specimens after 1 year of slide storage.
In preliminary studies using pig liver specimens, DNA yields were higher but DIN was lower when specimens were fixed in 20% rather than 10% formalin and when fixed for 3 or 7 days versus 1 day; however, the significance was dependent on specimen size and other preanalytical factors. In the same pig liver specimens, RNA yield and integrity decreased with prolonged fixation (3 or 7 days versus 1 day) although the magnitude and timing of effects was influenced by specimen size and formalin concentration. Both the yield and integrity of DNA and RNA isolated from pig liver were dependent on specimen size, with 2 mm needle specimens yielding the highest DNA yield and integrity, while RNA yields were highest from large specimens (50 mm thick) and RNA integrity was greatest from the smallest specimens (10 mm thick). Additional effects of RNA and DNA yield and integrity were associated with sampling location (from the surface versus the center of the specimen) but the significance of these differences depended on fixation duration and fixative concentration. A delay to fixation , had little to no effect on DNA or RNA yield or integrity in pig liver specimens, but delayed fixation of pig pancreas specimens led to significant declines in DNA and RNA integrity.
Biospecimens
TissueColorectalTissueUterusTissueOvaryTissuePancreasTissueLiverPreservation Types
FormalinDiagnoses
Not specifiedPlatforms
DNAAutomated electrophoresis/BioanalyzerDNAReal-time qPCRDNAFluorometryRNAAutomated electrophoresis/BioanalyzerRNASpectrophotometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration FFPE blocks stored <1 year versus > 1 yearSections stored for 0 days, 1 week, 1 month, 3 months, and 1 yearStorage Storage temperature -30°C-80°C4°CRoom temperature (20°C–22°C)Automated electrophoresis/Bioanalyzer Specific Technology platform DIN versus fragment length and ΔΔCqRIN versus DV200