This paper evaluated how the temperature and duration of storage of formalin-fixed, paraffin-embedded (FFPE) blocks affect the quality and amplification of isolated DNA and RNA. DNA and RNA were extracted from four case-matched FFPE blocks stored at 18°C, 4°C, -20°C, -80°C and -150°C for up to 12 months before sectioning.
Effect of storage temperature on nucleic acid quality in formalin-fixed paraffin-embedded tissue samples.
Okubo Y, Yokose T, Kasajima R, Saji H, Ito H, Kaburaki Y, Hasegawa C, Yuguchi S, Yoshioka E, Washimi K, Sato S, Hiroshima Y, Hayashi H, Miyagi Y, Toyama N
Purpose
Conclusion
When FFPE blocks were stored at 18°C, storage duration was negatively correlated with DNA integrity number (DIN; r=-0.655, P<0.05), the ratio of cycle threshold (CT) values for the 87 and 256 bp amplicons (SL Ct ratio; r=-0.857, P<0.05), and RNA integrity number (RIN; r=-0.685, P <0.05). Additionally, DV200 values decreased from 72.5% in specimens stored for 3 months to 51.6% in specimens stored for 12 months, but the difference was not statistically significant. When FFPE blocks were stored at 4°C, there was a negative correlation between storage duration and DIN (r=-0.661, P<0.05) and DV200 (r=-0.667, P<0.05), but not the SL Ct ratio or RIN. While small differences in DIN and DV200 were noted with frozen storage, they were not significant.
Studies (1)
Purpose
This study evaluated how the temperature and duration of storage of formalin-fixed, paraffin-embedded (FFPE) blocks affect the quality and amplification of isolated DNA and RNA. Surgically resected tumor specimens were collected from four patients: a patient with lung adenocarcinoma (44 year-old man), a patient with squamous cell carcinoma (68 year-old-man), a patient with endometrioid carcinoma (67 year-old-woman), and a patient with a seromucinous borderline tumor (87 year-old-woman). Specimens experienced cold ischemia times that ranged from 29 to 60 min before being divided into five parts and fixed in 10% neutral buffered formalin for 16-23 h. Blocks were paraffin embedded (details not provided). Matched FFPE blocks were stored at 18°C, 4°C, -20°C, -80°C and -150°C for 3, 6 and 12 months. Prior to sectioning, refrigerated/frozen blocks were placed in a container with desiccant and returned to room temperature. DNA was extracted from sections using the QIAamp DNA FFPE Advanced UNG Kit. DNA was quantified using a NanoDrop spectrophotometer and a Qubit Fluorometer. Fragment size was analyzed using a Genomic DNA ScreenTape assay and by real-time PCR amplification of a 87 bp RNase P amplicon and a 256 bp MGB amplicon. RNA was extracted using the RNeasy FFPE Kit. RNA was quantified using a NanoDrop spectrophotometer and the Qubit RNA BR Assay Kit. Fragment size was analyzed using a High Sensitivity RNA ScreenTape assay.
Summary of Findings
When FFPE blocks were stored at 18°C, storage duration was negatively correlated with DNA integrity number (DIN; r=-0.655, P<0.05), the ratio of cycle threshold (CT) values for the 87 and 256 bp amplicons (SL Ct ratio; r=-0.857, P<0.05), and RNA integrity number (RIN; r=-0.685, P <0.05). Additionally, DV200 values decreased from 72.5% in specimens stored for 3 months to 51.6% in specimens stored for 12 months, but the difference was not statistically significant. When FFPE blocks were stored at 4°C, there was a negative correlation between storage duration and DIN (r=-0.661, P<0.05) and DV200 (r=-0.667, P<0.05), but not the SL Ct ratio or RIN. While small differences in DIN and DV200 were noted with frozen storage, they were not significant.
Biospecimens
TissueOvaryTissueLungPreservation Types
FormalinDiagnoses
NeoplasticCarcinomaPlatforms
DNAFluorometryDNAReal-time qPCRDNAAutomated electrophoresis/BioanalyzerRNAAutomated electrophoresis/BioanalyzerRNAFluorometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 12 months3 months6 monthsStorage Storage temperature -150°C-20°C-80°C18°C4°C