Measurement of uric acid in FC Mix tubes is not accurate.

Reeve JL, O'Meara MJ, Griffin DG

Publication: Ann Clin Biochem (2024) • Page 45632241308076PMID: 39670292View on PubMed

Purpose

This paper compared uric acid levels in twenty-four case-matched specimens collected in Greiner FC Mix tubes (plasma) and Greiner serum separator tubes using two different clinical chemistry analyzers. All specimens were selected retrospectively and stored 1-3 days at 4°C before analysis.

Conclusion

Using the Roche Cobas instrument, uric acid levels differed significantly between the two specimen types (Wilcoxen P<0.0001), with higher levels observed in serum (from serum separator tubes) than matched plasma (from FC Mix tubes) in 22 of 24 pairs. Similarly, in all ten sample pairs analyzed with the Cobas instrument, serum levels were higher than matched plasma specimens. The authors conclude that FC Mix tubes are unsuitable for uric acid quantification in plasma.

Studies (1)

Purpose

This study compared uric acid levels in twenty-four case-matched specimens collected in Greiner FC Mix tubes (plasma) and Greiner serum separator tubes using two different clinical chemistry analyzers. All specimens were selected retrospectively and stored 1-3 days at 4°C before analysis.  Serum/plasma from 24 paired blood specimens collected in Greiner FC Mix (plasma) and Greiner serum separator tubes without interfering levels of lipemia, icterus or hemolysis were selected at random. All specimens were ordered by primary care providers for routine analysis (no diagnoses specified). Specimens had been centrifuged on the day of collection (conditions not specified) and stored 1-3 days at 4°C. Uric acid levels were quantified by an enzymatic colorimetric uric acid assay on a Roche Cobas c 702 instrument. Uric acid levels were analyzed in 10 of the 24 pairs by an enzymatic colorimetric uric acid assay and an Abbott Architect C4000 instrument.

Summary of Findings

Using the Roche Cobas instrument, uric acid levels differed significantly between the two case-matched plasma and serum specimens (Wilcoxen P<0.0001), with higher levels observed in serum (from serum separator tubes) than matched plasma (from FC Mix tubes) in 22 of 24 pairs. Similarly, in all ten sample pairs analyzed with the Cobas instrument, serum levels were higher than in the matched plasma specimen. The authors concluded that FC Mix tubes are unsuitable for uric acid quantification in plasma.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Greiner FC Mix
    Greiner serum separator tubes
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Clinical chemistry/auto analyzer SpecificTechnology platform
    Abbott Architect C4000 assay
    Roche Cobas c 702 assay