This paper investigated the effects of incomplete tube filling (25%, 50% and 100%) on the concentration and purity of RNA and DNA from K2EDTA blood collected from healthy volunteers. RNA and DNA concentrations were analyzed by spectrophotometry and by real-time PCR amplification of Abelson murine leukemia (ABL) and glyceralde-hyde-3-phosphate dehydrogenase (GAPDH), respectively.
Preanalytical Impact of Incomplete K(2)EDTA Blood Tube Filling in Molecular Biology Testing.
Benati M, Pighi L, Paviati E, Visconti S, Lippi G, Salvagno GL
Purpose
Conclusion
The concentration of RNA was significantly higher in the tube with only 1.5 mL (25% fill) or 3 mL (50% fill) versus 6 mL (100% fill) of blood (62.44 ng/μL and 59.84 ng/μL versus 56.58; p=0.001 and p=0.021, respectively). The copy number of ABL was also higher in RNA extracted from the partially filled tubes than the completely filled tubes, but the difference was only significant between the tube filled with 1.5 mL (25% fill) relative to the tube with 6 mL (100%) of blood (mean 8,225 copies versus 7,343 copies, p=0.026). While DNA concentration was not significantly affected by fill volume, the cycle threshold (CT) values for GAPDH were higher when the tube was filled with 1.5 mL (25% fill) or 3 mL (50% fill) versus 6 mL (100% fill) of blood (p=0.026 and p=0.017, respectively). RNA purity, as evaluated by the ratio of absorbance at 260/280 nm or 260/230 nm, was not affected by fill volume, but DNA purity, as measured by the 260/280 ratio, was higher when extraction was from 6 mL rather than 1.5 or 3 mL of blood (mean 1.75 both, versus 1.77, p=0.047 and p=0.021, respectively).
Studies (1)
Purpose
This study investigated the effects of incomplete tube filling (25%, 50% and 100%) on the concentration and purity of RNA and DNA from K2EDTA blood collected from healthy volunteers. Blood was collected from thirteen healthy volunteers (4 males and 9 females; 26-66 years of age) into a syringe using a 21-gauge needle. A total of 1.5 mL, 3 mL and 6 mL of blood were added to 6 mL K2EDTA vacutainers, resulting in a 25%, 50% and 100% fill volume, respectively. RNA was extracted from 1 mL of each blood specimen using the Maxwell RSC RNA Blood Kit and DNA was extracted from an unspecified volume of blood using the MagCorePlus II Kit. RNA and DNA were quantified and their purity evaluated by spectrophotometer. RNA and DNA were further quantified by real-time PCR of ABL and GAPDH, respectively.
Summary of Findings
The concentration of RNA was significantly higher in the tube with only 1.5 mL (25% fill) or 3 mL (50% fill) versus 6 mL (100% fill) of blood (62.44 ng/μL and 59.84 ng/μL versus 56.58; p=0.001 and p=0.021, respectively). The copy number of ABL was also higher in RNA extracted from the partially filled tubes than the completely filled tubes, but the difference was only significant between the tube filled with 1.5 mL (25% fill) relative to the tube with 6 mL (100%) of blood (mean 8,225 copies versus 7,343 copies, p=0.026). While DNA concentration was not significantly affected by fill volume, the CT values for GAPDH were higher when the tube was filled with 1.5 mL (25% fill) or 3 mL (50% fill) versus 6 mL (100% fill) of blood (p=0.026 and p=0.017, respectively). RNA purity, as evaluated by the ratio of absorbance at 260/280 nm or 260/230 nm, was not affected by fill volume but DNA purities, as measured by the 260/280 ratio, was higher when extraction was from 6 mL rather than 1.5 or 3 mL of blood (mean 1.75 both versus 1.77, p=0.047 and p=0.021, respectively).
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
DNASpectrophotometryDNAReal-time qPCRRNASpectrophotometryRNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Improperly filled tubeProperly filled tubeBiospecimen Aliquots and Components Aliquot size/volume 1.5 mL (25% filled)3 mL (50% filled)6 mL (100% filled)