Acceptability of plasma ammonia results when samples are not transported and processed under ideal conditions.

Bowron A, Osgood V

Publication: Ann Clin Biochem (2024) • Page 45632241232931PMID: 38298138View on PubMed

Purpose

This paper compared plasma ammonia levels in blood that was stored at room temperature for up to 2 h prior to centrifugation with levels in matched specimens that were stored on ice for 15 min prior to centrifugation.

Conclusion

Plasma ammonia levels were significantly higher in the specimens stored at room temperature for <15 min than in matched specimens that were stored on ice for 15 min prior to centrifugation and analysis. Ammonia levels increased linearly with storage of blood at room temperature, reaching 30±8.4 μmol/L after 120 min. Despite the large increase, levels remained below the reference limit of 50 μmol/L, leading the authors to conclude that ammonia may be analyzed in plasma isolated from blood specimens that are stored for up to 2 h at room temperature.

Studies (1)

Purpose

This study compared ammonia levels in plasma isolated from blood that was stored at room temperature for up to 2 h with levels in matched specimens from blood stored on ice for 15 min. Blood was collected from 11 healthy adult volunteers into syringes and then transferred into 14 K2EDTA Microvette tubes. One Microvette tube from each volunteer was immediately placed on ice for 15 min, whereas the remaining Microvettes were stored at room temperature. One of the tubes stored at room temperature from each volunteer was analyzed within 15 min (baseline), and subsequently an additional tube was analyzed every 10 min for 2 h. Plasma was isolated by centrifugation at 2000 g for 5 min at 4°C. Ammonia was immediately analyzed in plasma using a Roche c702 analyzer.

Summary of Findings

Plasma ammonia levels were significantly higher in the specimens stored at room temperature for <15 min than in matched specimens that were stored on ice for 15 min prior to centrifugation and analysis (19±6.6 μmol/L versus 18±6.6 μmol/L, P=0.02). Ammonia levels increased linearly with storage of blood at room temperature (R2 = 0.9896, P<0.001), reaching 30±8.4 μmol/L after an additional 120 min. Importantly, despite the large increase, levels in all specimens remained below the reference limit of 50 μmol/L. The authors concluded that ammonia may be analyzed in plasma from blood specimens that are stored for up to 2 h at room temperature.

 

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage temperature
    At room temperature
    On ice
    StorageTime at room temperature
    <15 min
    Additional 10 min
    Additional 100 min
    Additional 110 min
    Additional 120 min
    Additional 20 min
    Additional 30 min
    Additional 40 min
    Additional 50 min
    Additional 60 min
    Additional 70 min
    Additional 80 min
    Additional 90 min