Technical assessment of different extraction methods and transcriptome profiling of RNA isolated from small volumes of blood.

Kalikiri MKR, Manjunath HS, Vempalli FR, Mathew LS, Liu L, Wang L, Wang G, Wang K, Soloviov O, Lorenz S, Tomei S

Publication: Sci Rep (2023) • Vol 13 • Page 3598PMID: 36869090View on PubMed

Purpose

This paper compared the yield, purity, integrity, and transcriptome profile of RNA extracted from 50 µL of blood using a manual method (Tempus Spin RNA Isolation Kit) with and without a TURBO DNA Free treatment (TURBO) step and when RNA was isolated from 16-66 µL of blood using a partially automated method (MagMAX for Stabilized Blood RNA Isolation Kit).

Conclusion

The manual extraction method yielded a higher concentration of RNA and RNA with a higher RNA integrity number (RIN) than the partially automated method. When a TURBO DNA Free treatment (TURBO) step was performed in conjunction with the manual isolation method, RNA yield was lower, the average OD 260/280 nm ratio was < 2, and mean RIN was higher compared to when the TURBO step was not included. The concentration of isolated RNA increased when larger volumes of blood were used for extraction, although the variability in RNA concentration also increased. Manually extracted specimens had a slightly higher percentage of reads that mapped to exons but also more variable variance stabilizing transformation (VST) counts than those extracted using the partially automated method. However, overall coverage, alignment scores, and gene types were similar between the manual (TURBO treated or untreated), and partially automated methods; notably, specimens clustered by patient and not extraction method during principal component analysis. Further correlation analysis revealed strong correlations between specimens collected from the same patient regardless of extraction method, whether samples did or did not undergo TURBO  treatment, and the blood volume used during extraction (16-66 µL).

Studies (1)

Purpose

This study compared the yield, purity, integrity, and transcriptome profile of RNA extracted from 50 µL of blood using a manual method (Tempus Spin RNA Isolation Kit) with and without a TURBO DNA Free treatment (TURBO) step and when RNA was isolated from 16-66 µL blood using an partially automated method. Blood was collected from three healthy volunteers into capillary tubes.  RNA was isolated from 50 µL of blood using the manual Tempus Spin RNA Isolation Kit with and without a TURBO DNA Free Treatment step and from 16, 33, 50, and 66 µL aliquots of blood using a partially automated in-house modification (addition of manual steps to pellet and wash RNA) of the MagMAX for Stabilized Blood RNA Isolation Kit with and without the TURBO DNA Free treatment. RNA was quantified with either a NanoDrop 8000 Spectrophotometer or a Qubit RNA HS Assay Kit. RNA profile and integrity were then assessed using the RNA Assay Reagent Kit on the LabChip GXI instrument. Sequencing libraries were constructed with the Lexogen QuantSeq3′ mRNA-Seq libraries for Illumina Sequencing Kit and sequenced on an Illumina NextSeq 500 system.

Summary of Findings

Both RNA concentration and variability in concentration were higher when larger blood input volumes were used for extraction, although RIN was unaffected. When RNA was extracted from 50 µL of blood, mean RNA concentration was higher using the manual method without the TURBO DNA Free step compared to when the manual or partially automated method was used with a TURBO step (P=0.0010 and P=0.0129, respectively). RINs were higher when the TURBO treatment step was performed in conjunction with manual extraction than when it was excluded (P<0.0001), and when RNA was extracted using the manual method without the TURBO treatment step than when the partially automated method with TURBO was used (P<0.0001). RNA concentrations were significantly correlated between TURBO-treated and untreated extractions when extraction was with the manual method (R2=0.6294, P=0.001), but were not correlated when extraction was with the partially automated method. Interestingly, extraction using the manual method with the TURBO step was the only method that resulted in RNA with an OD 260/280 nm ratio < 2 (less pure).

Unacceptable or suboptimal libraries were generated from four of the 60 RNA extracts. Two manually extracted specimens from a single volunteer had unacceptable library yield and size which the authors attributed to a A260/A230 ratio that was “high”. Two other specimens generated libraries of suboptimal molarity and were described as low-concentration libraries.  Manually extracted specimens had a slightly higher percentage of reads that mapped to exons, but also more variable variance stabilizing transformation (VST) counts than those extracted using the partially automated method. However, overall coverage, alignment scores, and gene types were similar between manual (TURBO treated or untreated), and automated RNA extraction methods; further, specimens clustered by patient not extraction method during principal component analysis. Further correlation analysis revealed strong correlations between specimens from the same patient regardless of extraction method, inclusion of a TURBO step, and the volume of blood used (16-66 µL).

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    RNAAutomated electrophoresis/Bioanalyzer
    RNAFluorometry
    RNANext generation sequencing
    RNASpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    16 µL
    33 µL
    50 µL
    66 µL
    Analyte Extraction and PurificationAnalyte isolation method
    MagMAX for Stabilized Blood RNA Isolation Kit (automated)
    Tempus Spin RNA isolation Kit (manual)
    Analyte Extraction and PurificationNucleic acid digestion
    No TURBO DNA Free treatment step
    TURBO DNA Free treatment step included