Method Validation for Extraction of Nucleic Acids from Peripheral Whole Blood.

Mathay C, Hamot G, Henry E, Mommaerts K, Thorlaksdottir A, Trouet J, Betsou F

Publication: Biopreserv Biobank (2016) • Vol 14 • Page 520-529PMID: 27548617View on PubMed

Purpose

 This paper sought to optimize the extraction of DNA from blood and buffy coat specimens and RNA from PAXgene blood specimens by comparing yield and purity of DNA and RNA and RNA integrity. The optimization strategies assessed for DNA extraction included different volumes of blood (2, 3, and 4 mL), manual versus automated lysis, different elution volumes (300 or 600 µL), storage of lysate (not stored versus storage at 2- 8°C for 1 week), and storage of extracted DNA (in 13 mL and 1.5 mL tubes on Chemagenic racks or in 1.5 mL tubes not on racks). Optimization strategies for RNA extraction included different elution buffers (water, tris EDTA, and BR5 buffer), storage of blood (18-24 h at room temperature versus 1-2 weeks at 2-8°C or 2 weeks at -20°C), and extraction using different automated instruments (Biorobot, QiaCube, QiaSymphony and MSM).

Conclusion

DNA yield and purity (determined by OD A260/A280) were comparable in specimens with different starting blood volumes (2 mL, 3 mL, or 4 mL), elution volumes (300 or 600 µL), and lysis methods (manual versus automated). Consequently, the authors chose to use 4 mL of blood for DNA extraction by manual lysis with an elution volume of 300 µL to investigate reproducibility. The coefficient of variance (CV) of DNA yield and purity were <10% in replicate samples that were extracted from the same lysate extracted on three consecutive days. The method was considered robust as the yield and purity of DNA was acceptable (>15 µg DNA, A260/A80 >1.8 and CV <10%) regardless of whether blood or buffy coat was used and whether lysate was stored for 1 week at 2-8°C prior to extraction, and whether isolated DNA was stored with or without a magnetic rack.

RNA yield was lower when blood was stored for one week at -20°C prior to extraction; RNA integrity numbers (RIN) were higher when blood was stored at 2–8°C for 2 weeks or at -20°C for 1 week than when blood was stored for 18-24 h at room temperature but yield and purity were comparable when RNA was eluted in water, tris EDTA or BR5 elution buffer. Isolation of RNA with the automated extraction MSM instrument resulted in low RNA purity (A260/A280 =1.52). RNA extracted with a Biorobot had a low mean RIN (6.9), but all other instruments yielded RNA of acceptable purity (A260/A280 >1.8) and integrity (mean RIN >7). Spectrophotometric RNA yields were higher when the QIAsymphony as used for extraction as opposed to the Biorobot instrument. Levels of miR-16 were highest when RNA extraction was with the Biorobot, QiaCube and QiaSymphony instruments and levels of RNU24 were higher when extraction was with QiaSymphony instrument in comparison to extraction with the MSM or Biorobot. When the protocol used with the Biorobot was changed to “No clot detection” mean RNA yields (6.57µg) and purity (A260/A280 =1.95) were higher than using the initial program; nevertheless, yields were slightly higher with the QiaCube than the optimized Biorobot protocol.  Specimens that were considered “difficult to lyse”, had mean RNA yields that were comparable when either the QiaCube or the Biorobot instrument was used, and the mean RIN and purity were acceptable when extraction was with either instrument.

Studies (2)

Purpose

This study sought to optimize the extraction of DNA from blood and buffy coat specimens by comparing DNA yield and purity following extraction from different volumes of blood (2, 3, and 4 mL), using manual versus automated lysis, using different elution volumes (300 or 600 µL), and following storage of lysate (not stored versus 2- 8°C  for 1 week) and storage of extracted DNA(in 13 and 1.5 mL tubes on Chemagenic racks or 1.5 mL tubes not on racks).  Blood was collected from three or more volunteers (no diagnosis) into K2EDTA vacutainer tubes and centrifuged at 2,000 x g for 20 min at 18-22°C. Buffy coat was automatically removed using a Freedom EVO liquid handler and then 1 mL aliquots were transferred to conical tubes and diluted with 4 mL water.  DNA was extracted from blood and buffy coat using the Chemagic DNA Blood Kit Kit and the automated Chemagic MSM I protocol. Extracted DNA was stored at -80°C. DNA yield and A260/A280 were assessed by spectrophotometer. To Potential effects of the volume of blood used for DNA extraction was investigated using specimens collected from three healthy volunteers; DNA was extracted from 2, 3, and 4 mL aliquots of blood diluted with water to obtain a final volume of 4 mL. To investigate the Potential effects of manual and automated lysis were investigated using the blood and isolated buffy coat of two volunteers; buffy coat specimens were lysed manually with a vortex, digested with proteinase K and used immediately for extraction while matched aliquots were treated with lysis buffer with proteinase K using the non-prefilled program on the MSM I instrument and then stored at room temperature for 2 h before extraction. Potential effects of elution volume were investigated using blood collected from two volunteers; blood specimens were manually lysed and DNA was extracted using a MSM1 instrument and eluted with either 200 or 600 µL of elution buffer. Reproducibility was assessed using blood aliquots from a single donor; blood aliquots were lysed and stored at 2-8°C before extraction on three consecutive days.  Robustness of the DNA extraction method was investigated by extracting DNA from matched blood and buffy coat from a single volunteer immediately or after lysate was stored for 1 week at 2-8°C.  The robustness of storing extracted DNA on a Chemagen magnetic rack was investigated using blood collected from two volunteers; DNA extracted from buffy coat stored in 13 mL and 1.5 mL tubes in Chemagen magnetic racks compared to matched DNA samples stored in 1.5 mL tubes that were not placed on the rack.

Summary of Findings

DNA yield and purity (A260/A280) were not significantly different (based on acceptance criteria) when DNA was extracted from 2 or 3 mL of blood instead of 4 mL; further,  the DNA yield from 2 and 3 mL of blood was greater than the acceptable threshold for 4 mL of blood (>40% from 2 mL and >70% from 3 mL). DNA yield and purity also met the acceptance criteria (no significant difference in yield or A260/A280 ratio) regardless of the elution volume (300 or 600 µL), or lysis method (manual versus automated).  Consequently, the authors opted to use 4 mL of blood, manual lysis, and a 300 µL elution volume to investigate reproducibility. The coefficient of variance (CV) of DNA yield and purity were <10% in replicate DNA extractions from the same lysate conducted on three consecutive days. The method was considered robust, as the yield and purity of DNA was acceptable (>15 µg DNA, A260/A80 >1.8, and CV <10%) regardless of whether blood buffy coat was used for extraction, whether the lysate was or stored for 1 week at 2- 8°C prior to extraction or extraction was immediate, and whether isolated DNA was stored at -80°C with or without a magnetic rack.

 

  • Biospecimens

    FluidBuffy Coat
    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    DNASpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    2 mL
    3 mL
    4 mL
    Biospecimen Aliquots and ComponentsBlood and blood products
    Buffy coat
    Whole blood
    StorageStorage conditions
    Extracted DNA not on magnetic rack in 1.5 mL tube
    Extracted DNA on magnetic rack in 1.5 mL tube
    Extracted DNA on magnetic rack in 13 mL tube
    StorageStorage duration
    0 h
    1 week
    Analyte Extraction and PurificationAnalyte isolation method
    Automated lysis
    Manual lysis
    Analyte Extraction and PurificationRehydration of dried sample/specimen
    300 µL
    600 µL