Skip to main content

Evaluation of a pneumatic tube system carrier prototype with fixing mechanism allowing for automated unloading.

Märtens CM, Schöpfel J, Bollmann S, Hannemann A, Zylla S, Dahl MB, Gauß F, Schedl J, Nauck M, Petersmann A

Publication: Clin Chem Lab Med (2022) • Vol 60 • Page 1202-1210PMID: 35635785View on PubMed

Purpose

The purpose of this paper was to compare levels of 51 analytes in blood specimens transported by courier, and by pneumatic tube system (PTS) using both a conventional carrier and a prototype carrier that automatically fixes the tubes in place.  

Conclusion

Of the 51 analytes evaluated, only levels of free hemoglobin and lactate dehydrogenase (LDH) were significantly impacted by transportation method, and for differences in median levels of were smaller than measured imprecision for both analytes. Levels of each analyte were generally strongly correlated among specimens transported using different transportation methods (r=0.75-1.0, median 0.96-0.97).

Studies (1)

Purpose

The purpose of this study was to compare levels of 51 analytes in blood specimens transported by courier, and by PTS using both a conventional carrier and a prototype carrier that automatically fixes the tubes in place.  Venous blood was collected from 35 healthy volunteers into triplicate lithium heparin BD Vacutainer PST II tubes with an inert polymer separator gel, K2EDTA BD Vacutainer, and sodium citrate BD Vacutainer tubes. Matched blood specimens were transported by courier, and by PTS using conventional and prototype carriers. The prototype carrier holds nine tubes that are automatically fixed in place by the lid. After arrival in the laboratory the lithium heparin (used for clinical chemistry determinations) and sodium citrate tubes (used for coagulation assays) were centrifuged at 3280 g for 5 min. A total of 51 analytes were quantified using autoanalyzers. The Friedman test was used to determine if analyte levels were significantly affected by transportation method. The Wilcoxon singed-rank test was applied to evaluate the significance of differences in analyte levels between each pair of transportation methods.

Summary of Findings

Of the 51 analytes, only free hemoglobin and LDH levels were significantly impacted by transportation method. Free hemoglobin levels were higher in specimens transported by PTS than those transported by courier (P<0.01 for conventional carrier and P<0.001 for the prototype carrier), and in specimens transported by PTS using the prototype carrier rather than the conventional carrier (P<0.05).

LDH levels were significantly higher in specimens transported by PTS than by courier (P<0.05, both carrier types), but LDH levels were not significantly affected by the type of PTS carrier. Importantly, differences in median levels of free hemoglobin and LDH between transportation methods were smaller than the measured imprecision. Levels of the analytes evaluated were strongly to very strongly correlated between specimens transported by courier and PTS with a conventional carrier (r=0.77-1.0, median r= 0.96 ) and between those transported by courier and PTS with the prototype carrier (r=0.75-1.0, median r= 0.97).  Blood transportation method did not affect levels of aTT, ASAT, absolute basophil, relative basophil, chloride, cholesterol, creatinine, absolute  eosinophil, relative eosinophils, erythrocyte, absolute erythrocyte volume distribution, Factor VII, fibrinogen, fT3, fT4, GGT, glucose, hepatoglobin, HDL cholesterol, hematocrit, hemoglobin, hemoplexin, LDL cholesterol leukocyte lipase, absolute lymphocyte, relative lymphocyte, magnesium, mean corpuscular hemoglobin, MCHC, MCV, absolute monocytes, relative monocytes, MPV, absolute neutrophils, relative neutrophils, phosphate, platelet count, potassium,  protein, sodium, prothrombin time, thrombin time, thrombocyte distribution width, triglycerides, TSH, or uric acid were identified.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Cell count/volumeHematology/ auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    LipidClinical chemistry/auto analyzer
    MorphologyHematology/ auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    ProteinHematology/ auto analyzer
    Small moleculeClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageSpecimen transport duration/condition
    Prototype carrier that automatically fixes tubes in place
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system