Effects of various pre-analytical conditions on blood-based biomarkers of Alzheimer's disease.

Ladang A, Rigaud L, Sqalli G, Cavalier E

Publication: Clin Chem Lab Med (2021) • Vol 59 • Page e435-e437PMID: 34233385View on PubMed

Purpose

This paper compared levels of amyloid-β (Aβ) and total Tau (t-Tau) in plasma from blood of healthy men collected in K2EDTA Vacutainer, K3EDTA S-Monovette, and K2EDTA Microtainer MAP tubes that were subjected to 4 or 8 tube inversions and centrifuged immediately or after a 1 h delay in a rack or lying on the bench.

Conclusion

Aβ1-42, and Aβ1-40 levels were significantly higher when blood was collected in a K2EDTA Vacutainer rather than a K2EDTA Microtainer tube, but the ratio of Aβ1-42 to Aβ1-40 was unaffected by the type of blood collection tube. Levels of Aβ1-42, Aβ1-40 and t-Tau in plasma from K3EDTA S-Monovette and K2EDTA Microtainer tubes were comparable among the four processing conditions, but Aβ1-40 levels were higher in specimens collected in K2EDTA vacutainer tubes that were left lying on the bench for an hour before centrifugation than in those that were processed immediately.

Studies (1)

Purpose

This study compared levels of Aβ1-42, Aβ1-40 and t-Tau in plasma from blood of healthy men collected in K2EDTA Vacutainer, K3EDTA S-Monovette, and K2EDTA Microtainer MAP tubes that were subjected to 4 or 8 inversions and centrifuged immediately or after a 1 h delay (at room temperature) in a rack or lying on the bench. Blood was collected from seven healthy men (aged 20-35 years) into K2EDTA Vacutainer, K3EDTA S-Monovette, and K2EDTA Microtainer MAP tubes. One tube of each type was subjected to each of the following processing: i.) inverted 8 times, centrifuged (2860 g for 5 min) and aliquoted immediately; ii.) inverted 4 times, centrifuged (2860 g for 5 min), and aliquoted immediately; iii.) inverted 8 times, stored in rack at room temperature for 1 h, centrifuged (2860 g for 5 min)  and aliquoted; and iv.) inverted 8 times, stored lying on the bench at room temperature for 1 h, centrifuged (2860 g for 5 min) and aliquoted. Plasma aliquots were stored frozen at -80°C for less than 15 days before analysis. Levels of Aβ1-42, Aβ1-40 and t-Tau in plasma were quantified with SiMoA technology using Neurology 3-Plex A on a SR-X platform analyzer.  For each specimen, levels were quantified in two runs with one freeze-thaw cycle between runs.

Summary of Findings

Mean Aβ1-42, Aβ1-40 and t-Tau levels were 15%, 13% and 16% higher, respectively, when blood was collected in K2EDTA Vacutainer rather than K2EDTA microtainer tubes, but the difference was only significant for Aβ1-42 (P=0.016) and Aβ1-40 (P=0.031).  The ratio of Aβ1-42 to Aβ1-40 was unaffected by the type of blood collection tube. Levels of Aβ1-42, Aβ1-40 and t-Tau in plasma from K3EDTA S-Monovette and K2EDTA Microtainer tubes were comparable among the four processing conditions evaluated, but Aβ1-40 was 14-15% higher in specimens from K2EDTA Vacutainer tubes that were left lying on the bench for an hour (protocol iv) compared to those that were processed immediately (protocol i or ii).

 

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    ProteinClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    K2EDTA Microtainer MAP tubes
    K2EDTA Vacutainer
    K3EDTA S-Monovette
    Biospecimen Aliquots and ComponentsBiospecimen mixing
    Inverted 4 times
    Inverted 8 times
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage conditions
    In rack
    Lying on bench
    StorageTime at room temperature
    0 h
    1 h