This paper compared the stability of glucose in EDTA plasma, sodium fluoride (NaF) plasma and serum subjected to pre-centrifugation or post-separation storage at room temperature and in a cold box (4-15°C).
Alternative pre-analytic sample handling techniques for glucose measurement in the absence of fluoride tubes in low resource settings.
Nakanga WP, Balungi P, Niwaha AJ, Shields BM, Hughes P, Andrews RC, Mc Donald TJ, Nyirenda MJ, Hattersley AT
Purpose
Conclusion
Glucose levels were comparable in NaF plasma, EDTA plasma and serum when centrifugation was within 30 min of venipuncture and plasma/serum were immediately frozen. Storage of blood for ≥ 6 h at room temperature prior to centrifugation resulted in significantly lower glucose levels in the resultant serum and EDTA plasma than when centrifugation occurred within 30 min. However, glucose levels in NaF plasma were comparable among blood centrifuged within 30 min or after 24 h at room temperature. Pre-centrifugation storage of blood in a cold box (4-15°C) for 6 h resulted in significantly lower glucose levels in the resultant serum, but no significant changes in NaF or EDTA plasma. When blood was immediately centrifuged, glucose levels were unaffected by subsequent storage of serum, NaF plasma or EDTA plasma for 24 h in a cold box or at room temperature.
Studies (1)
Purpose
This study investigated potential effects of blood storage at room temperature and in a cold box prior to centrifugation on glucose levels in EDTA plasma, sodium fluoride (NaF) plasma and serum. Blood was collected with an 18-gauge needle from nine healthy volunteers (7 males and 2 females, aged 24-50 years) into EDTA, plain serum and sodium fluoride tubes. Plasma and serum were separated by centrifugation at 3000 g for 10 min within 30 min of venipuncture or after storage in a cold box (4-15°C) or at room temperature (25-28°C) for 0, 2, 6, 12, and 24 h. Plasma/serum aliquots were stored in a cold box or at room temperature for 0 (all processing delay specimens), 2, 6, 12, and 24 h before freezing at -20°C. Glucose levels were quantified using a Cobas 6000 analyzer.
Summary of Findings
Glucose levels were comparable in case-matched NaF plasma, EDTA plasma and serum when centrifugation was within 30 min of venipuncture and specimens were immediately frozen (5.4 mmol/L, 5.5 mmol/L and 5.2 mmol/L, respectively). When blood was stored in serum tubes at room temperature for 6 or 24 h prior to centrifugation, glucose levels in the resultant serum were only 64% and 6%, respectively, of levels in control serum that was isolated within 30 min (P=0.004, both). However, when a pre-centrifugation delay of 2 or 6 h occurred in a cold box (4-15°C),, then serum glucose levels were 99% and 92%, respectively, of levels in control serum that was isolated within 30 min (P=1 and P=0.004, respectively). Similarly, EDTA plasma separated from blood that was stored at room temperature for 6 or 24 h prior to centrifugation had glucose levels that were 77% and 34% of those in EDTA plasma that was isolated within 30 min (P=0.004, both), while glucose levels in EDTA plasma from blood stored for 6 h in a cold box were 97% of levels observed in control EDTA plasma isolated within 30 min. In contrast, a pre-centrifugation delay of 24 h at room temperature or 6 h in a cold box resulted in only small (6% and 0%) non-significant declines in glucose levels in NaF plasma. When blood was immediately centrifuged, glucose levels were unaffected by subsequent storage of serum, NaF plasma or EDTA plasma for 24 h in a cold box or at room temperature.
Biospecimens
FluidBloodFluidPlasmaFluidSerumPreservation Types
FrozenDiagnoses
NormalPlatforms
CarbohydrateClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Anticoagulant EDTANoneSodium fluorideBiospecimen Acquisition Type of collection container/solution EDTA tubeNaF tubeSerum tubeBiospecimen Aliquots and Components Blood and blood products PlasmaSerumWhole bloodBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage conditions As bloodAs plasma/serumStorage Storage duration 0 h12 h2 h24 h6 hStorage Storage temperature Cold boxRoom temperature