Metabolite Ratios as Quality Indicators for Pre-Analytical Variation in Serum and EDTA Plasma.

Heiling S, Knutti N, Scherr F, Geiger J, Weikert J, Rose M, Jahns R, Ceglarek U, Scherag A, Kiehntopf M

Publication: Metabolites (2021) • Vol 11PMID: 34564454View on PubMed

Purpose

The purpose of this paper was to investigate effects associated with delayed centrifugation of blood on levels of metabolites in serum and EDTA plasma from healthy patients. Markers of delayed centrifugation were then verified in the plasma and serum of patients with rheumatic or cardiovascular disease.

Conclusion

The effect of a delay to centrifugation of 2 h was larger in serum than EDTA plasma specimens, with 10.4% and 3.9% metabolites affected, respectively. Of the metabolites detected in serum, 59 increased and 11 decreased with a delay to centrifugation of 2 h compared to controls that were centrifuged within 30 min, but only 12 metabolites displayed changes >2-fold. In EDTA plasma, levels of 21 metabolites increased and four decreased with a 2 h delay to centrifugation but only one metabolite increased by >2-fold. Further examination of 8 metabolites showed a progressive increase in xanthine, hypoxanthine, and dihydroorotatic acid and a decrease in guanosine and inosine with delayed centrifugation of serum; while progressive increases in hypoxanthine and dihydroorotic acid were observed with delayed centrifugation of EDTA plasma. The ratios of xanthine and hypoxanthine to inosine or guanosine in serum also increased progressively with delayed centrifugation. Analysis of the eicosanoids identified revealed higher levels in serum than plasma.  While levels of most of the eicosanoids identified increased in serum when centrifugation was delayed by 4 h, in plasma that experienced the same delay only (±)9- hydroxyoctadecadienoic acid (HODE) and (±)13-HODE increased. When centrifugation was delayed serum and EDTA plasma concentrations of ornithine increased, arginine decreased, and the ornithine to arginine ratio increased. Among all the markers analyzed in serum, guanosine, hypoxanthine, 13-HODE, 15(S)-HETE, 12(S)-HEP, and 8(S)-HETE, 12-HETE and 12-oxo-ETE provided a high sensitivity (80-100%) and specificity (86-100%) for a >2 h delay to centrifugation and the ratio of hypoxanthine to inosine or provided the highest sensitivity (80-86%) and specificity (83-100%) for a delay to centrifugation of > 1 h. In plasma, the ratio of ornithine to arginine showed the highest sensitivity and specificity for a delay to centrifugation of 1 h or 2 h. In serum from patients with cardiovascular or rheumatic disease, the ratio of hypoxanthine to inosine or guanosine alone or together had a high degree of sensitivity (79-92%) and specificity (73-100%) for a delay to centrifugation of >1 h and several of the eicosanoids provided a high degree of sensitivity (92-100%) and specificity (86%) for a delay to centrifugation of >2 h. In plasma from patients with rheumatic or cardiovascular disease, only the hypoxanthine to inosine ratio was predictive of a delay to centrifugation of >1 h.

Studies (1)

Purpose

The purpose of this study was to investigate the effects of delayed centrifugation of blood on metabolite levels in serum and EDTA plasma from healthy patients. Markers of delayed centrifugation were then verified in the plasma and serum of patients with rheumatic or cardiovascular disease. Blood was collected using a 21-gauge needle from 43 healthy volunteers, 60 patients with cardiovascular disease, and 70 patients with a rheumatic disease was into Serum-gel-S Monovette tubes with a clot activator and K3EDTA tubes. Blood was stored in matched EDTA and serum tubes for 0.5, 1, 2, and 4 h before centrifugation at 2500 x g for 10 min at 20°C and at 3309 x g for 10 min at 20°C. After centrifugation, plasma and serum were stored at -80°C. The metabolite profile was analyzed in plasma and serum from 10 healthy volunteers with a 30 min and 2 h centrifugation delay by UPLC/MS with columns optimized for hydrophilic, hydrophobic, polar, and nonpolar compounds.  The effect of up to 4 h delayed centrifugation on nucleosides was investigated using MS-Omics in the serum and plasma of 10 healthy volunteers using MS-Omics. The effect of up to 4 h delayed centrifugation on eicosanoids was investigated using Shimadzu UFLC LC-20A Prominence liquid chromatography and a 5500 QTrap mass spectrometer in serum and plasma from ten healthy volunteers. The effect of up to 4 h delayed centrifugation on amino Acids was investigated in the serum and plasma of 23 healthy volunteers using a Biochrom 30+ Amino Acid Analyzer. Changes in metabolite levels with delayed centrifugation were verified by targeted MS/MS in the plasma and serum from patients with cardiovascular and rheumatoid disease.

Summary of Findings

More pronounced effects of a delay to centrifugation were observed in serum specimens than EDTA plasma (10.4% versus 3.9% of metabolites affected, respectively). Of the metabolites detected in serum, 59 increased and 11 decreased with delayed centrifugation of 2h compared to controls that were centrifuged after 30 min of venipuncture. An increase > 2-fold  in serum metabolite levels were observed for 7 metabolites after a 2 h delay while a decline > than 0.5-fold were observed for 5 metabolites. In EDTA plasma, levels of 21 metabolites increased and four decreased with a delayed in centrifugation of 2 h but only one metabolite increased by >2-fold. Metabolites in serum that were affected by a delay to centrifugation included saturated fatty acids, ethanolamines, linoleic acids, eicosanoids and dipeptide metabolic clusters and were involved in the glutamate, methionine, cysteine, S-adenosylmethionine (SAM), taurine, purine, lipid, amino acid, sphingolipid, arginine, ornithine metabolic pathways and the urea cycle. Metabolites in EDTA plasma that were affected by a delay to centrifugation included those involved in the endocannabinoids, fatty acid, arginine and ornithine metabolic pathways and urea cycle pathways. Further examination of eight metabolites (xanthine, hypoxanthine, inosine, guanosine, dihydroorotate, N-carbamoylaspartate, 7-methylguanine, and allantoin) chosen from clusters of metabolites in serum that were affected by delayed centrifugation, showed a progressive increase in xanthine, hypoxanthine, and dihydroorotatic acid and progressive decreases in guanosine and inosine with delayed centrifugation. In serum, the ratio of both xanthine and hypoxanthine to inosine or guanosine   also increased progressively with delayed centrifugation. In EDTA plasma only hypoxanthine and dihydroorotic acid were affected by a delay to centrifugation of up to 4 h. Further analysis of eicosanoids revealed higher levels in serum than plasma, increases in most eicosanoids in serum when centrifugation was delayed for 4 h, and increases in (±)9- hydroxyoctadecadienoic acid (HODE) and (±)13-HODE with delayed centrifugation of plasma.  Among the eicosanoids found in serum, 12- hydroxyeicosatetraenoic acid (HETE) was the best discriminator of delays to centrifugation. Ornithine concentrations increased and arginine concentrations decreased in both serum and EDTA plasma when centrifugation was delayed by 4 h.  The corresponding ornithine to arginine ratio increased progressively with delayed centrifugation in serum and plasma. Interestingly, ornithine concentrations were similar in plasma and serum obtained after 30 min, but arginine concentrations in EDTA plasma were only 65% of that in serum when centrifuged after 30 min. Of all the markers analyzed in serum, guanosine, hypoxanthine, 13-HODE, 15(S)-HETE, 12(S)-HEP, and 8(S)-HETE, 12-HETE and 12-oxo-ETE provided both high sensitivity (80-100%) and specificity (86-100%) for a >2 h delay to centrifugation, and the ratio of hypoxanthine to inosine provided the highest sensitivity (80-86%) and specificity (83-100%) for a delay to centrifugation of > 1h. In EDTA plasma, the ratio of ornithine to arginine showed the highest sensitivity and specificity for a delay to centrifugation of 1 h or 2 h. The ratio of hypoxanthine to inosine or guanosine had sensitivities of 79% and 100%, respectively, and specificities of 85% and 73%, respectively, for a time to centrifugation of >1 h in serum from patients with cardiovascular or rheumatic disease; and, in these patients the combination of markers had a sensitivity of 100% and specificity of 92%. Several of the eicosanoids provided a high degree of sensitivity (92-100%) and specificity (86%) for a delay to centrifugation of >2 h in serum. In EDTA plasma, only hypoxanthine to inosine ratio was predictive of a delay to centrifugation of >1 h in patients with rheumatic or cardiovascular disease.

  • Biospecimens

    FluidBlood
    FluidPlasma
    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
    Rheumatoid Arthritis
    Cardiovascular Disease
  • Platforms

    Small moleculeLC-TOF-MS
    Small moleculeMS/MS
    Small moleculeLC-MS or LC-MS/MS
    LipidLC-MS or LC-MS/MS
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBlood and blood products
    Plasma
    Serum
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageTime at room temperature
    1 h
    2 h
    30 min
    4 h