Effect of Pneumatic Tube System Transport on Cell-Free DNA.

Geerlings MJ, Hofste LSM, Kamping EJ, Abdi Z, Tolmeijer SH, Garms LM, Klarenbeek BR, Ligtenberg MJL

Publication: Clin Chem (2021) • Vol 67 • Page 434-435PMID: 33280007View on PubMed

Purpose

This paper compared two transportation methods (courier, pneumatic tube system (PTS)) regarding potential differences in the number of detectable copies of three different DNA amplicons in blood collected in Roche cell-free DNA tubes and EDTA tubes.

Conclusion

Transportation of EDTA blood by PTS resulted in more copies of all three-sized DNA amplicons in plasma compared to when transported by by courier; however, transportation method of blood collected in Roche cell-free DNA tubes did not affect copy numbers of differently sized DNA amplicons. The authors conclude that DNA is released from blood cells in EDTA tubes, but not Roche cell-free DNA tubes, during PTS transport.

Studies (1)

Purpose

This study compared two transportation methods (courier, PTS) regarding potential differences in the number of detectable copies of three different DNA amplicons in blood collected in Roche cell-free DNA tubes and EDTA tubes.  Blood was collected from a total of 25 patients with esophageal cancer into either duplicate EDTA tubes (8 patients) or Roche cell-free DNA blood collection tubes (17 patients). One specimen from each set was transported by PTS (840 m at 3.5-4.0 m/sec) and the other was transported by courier. Plasma was obtained by centrifugation at 1600 x g for 10 min followed by 16,000 x g for 10 min within 4 h (EDTA tubes) or 4 days (Roche cell-free DNA tubes) of collection. Plasma was stored at -80°C until DNA isolation using the QiaAmp Circulating Nucleic Acid Kit. DNA was quantified using a multiplex ddPCR assay that amplified 137, 420, and 1950 bp fragments of β-Actin.

Summary of Findings

Transportation of EDTA blood by PTS resulted in more copies of the 137, 420 and 1950 bp βActin amplicons in plasma compared to case-matched specimens transported by courier (+53%, P=0.002; +59%, P=0.004; and +111%, P=0.001, respectively). In contrast, transportation method of blood in Roche cell-free DNA tubes did not significantly affect copy numbers of the 137, 420 and 1950 bp amplicons (+3%, +3%, and +15%, respectively). The authors conclude that DNA is released from blood cells in EDTA tubes but not Roche cell-free DNA tubes during PTS transport.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    FrozenOther Preservative
  • Diagnoses

    NeoplasticCarcinoma
  • Platforms

    DNADigital PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    EDTA tube
    Roche cell-free DNA tubes
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system
    Digital PCR SpecificLength of gene fragment
    137 bp
    1950 bp
    420 bp
    Digital PCR SpecificTargeted nucleic acid
    β-actin