The effects of transport by pneumatic tube system on urine analysis.

Kosem A, Topcuoglu C, Sezer S, Demir E, Turhan T

Publication: Scand J Clin Lab Invest (2021) • Page 1-4PMID: 34346281View on PubMed

Purpose

The purpose of this paper was to investigate if pneumatic tube system (PTS) transport of urine alters erythrocyte, leukocyte, or epithelial cell counts or the presence of crystals, casts, or yeast cells compared to case--matched urine specimens transported manually.

Conclusion

The authors conclude that PTS transport does not affect urine analysis of erythrocytes, leukocytes, epithelial cells, crystals, casts, and/or yeast cells.

Studies (1)

Purpose

The purpose of this study was to compare erythrocyte, leukocyte, and epithelial cell counts and the presence of crystals, casts, and yeast cells in case-matched urine specimens transported manually or via PTS. Urine was self-collected from 187 outpatients (diagnosis not specified) and aliquoted into two tubes that were transported to the laboratory manually or by PTS. PTS involved transport for approximately 180 m at 6 m/s in a carrier with sponge rubber and plastic bags. Urine particles (cells, crystals, and casts) were identified, characterized and quantified using an IRIS iQ200 automated urine analyzer within 2 h of urine collection. Specimens were grouped based on cell count (0-5, 6-10, 10-30 and >30 cells per sample) prior to comparing effects of transport on erythrocytes, leukocytes and epithelial cells.

Summary of Findings

Erythrocyte, leukocyte and epithelial cell counts in PTS and manually transported urine specimens was very strongly associated (Gamma =0.982, 0.959, and 1.0, respectively). Similarly, the presence of crystals, casts and yeast cells in PTS and manually transported urine specimens were strongly associated (κ=0.952, 0.866, and 1.0, respectively). The authors conclude that PTS transport does not affect urine analysis of erythrocytes, leukocytes, epithelial cells, crystals, casts, and/or yeast cells.

  • Biospecimens

    FluidUrine
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageWithin hospital transportation method
    Hand-delivered
    Pneumatic tube system