This paper compared cycle threshold (CT) values for SARS-CoV-2, the virus that causes the novel 2019 coronavirus disease (COVID, COVID-19, COVID19), in respiratory specimens over the course of storage for up to 21 days at three temperatures and after up to 15 freeze/thaw cycles.
Prolonged Unfrozen Storage and Repeated Freeze-Thawing of SARS-CoV-2 Patient Samples Have Minor Effects on SARS-CoV-2 Detectability by RT-PCR.
Dzung A, Cheng P, Stoffel C, Tastanova A, Turko P, Levesque MP, Bosshard PP
Purpose
Conclusion
SARS-CoV-2 RNA was detected in all specimens, regardless of storage temperature or duration, but small changes in CT values (0.002-0.046 cycles per day) were noted. The change in CT value for each of the three genes was smaller when stored at 4°C than 35°C or 25°C. Larger changes in CT values were observed when specimens were freeze/thaw cycled five times compared to those only freeze/thaw cycled once and CT values increased further when subjected to ten freeze/thaw cycles.
Studies (1)
Purpose
This study compared CT values for three SARS-CoV-2 genes in respiratory specimens over the course of storage for up to 21 days at three temperatures and after up to 15 freeze-thaw cycles. The study included residual nasopharyngeal swab specimens collected into viral transport media (VTM) from five patients with SARS-CoV-2. Specimens were stored for <30 days at -20°C before they were thawed, diluted, and aliquoted into cryotubes for this study. Aliquots were stored at 4, 25, or 35°C for 1, 2, 3, 4, 5, 6, 7, 10, 14, or 21 days and then stored at -80°C or were frozen in liquid nitrogen and thawed at room temperature 1, 5, 10, or 15 times before storage. RNA was extracted from all specimens using the KingFisher Flex Purification System. SARS-CoV-2 ORF1ab, S gene, and N genes were detected using the real-time PCR TaqMan SARS-CoV-2 Assay Kit v2.
Summary of Findings
SARS-CoV-2 RNA was detected in all specimens, regardless of storage temperature or duration, but small changes in CT values (0.002-0.046 cycles per day) were noted. The change in CT value for each of the three genes was smaller when stored at 4°C (0.002-0.006 cycles/day) than 35°C (0.023-0.046 cycles/day, P<0.0001) or 25°C (0.008-0.036 cycles/day, P<0.0001). CT values were significantly higher in specimens subjected to five freeze/thaw cycles compared to those only freeze/thaw cycled once (P=0.0005). Specimens subjected to ten freeze/thaw cycles had higher CT values than those subjected to five cycles (P<0.0001) but CT values were comparable in specimens subjected to ten and 15 freeze/thaw cycles. The authors found for every freeze/thaw cycle, the average CT value of the N-gene, ORF1ab, and S genes increased by 0.197, 0.127, and 0.106 cycles, respectively.
Biospecimens
CellNasopharynxPreservation Types
FrozenDiagnoses
Pneumonia/Respiratory InfectionPlatforms
RNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Freeze/thaw cycling 1 cycle10 cycles15 cycles5 cyclesStorage Storage duration 0 days1 day10 days14 days2 days21 days3 days4 days5 days6 days7 daysStorage Storage temperature 25°C35°CReal-time qRT-PCR Specific Targeted nucleic acid N geneORF1abS gene