This paper investigated the effects of preanalytical factors including tube types, fill volume, mixing methods, storage duration, centrifugation, and transport conditions on the measurement of Alzheimer’s biomarkers in cerebrospinal fluid (CSF).
Pre-analytical protocol for measuring Alzheimer's disease biomarkers in fresh CSF.
Hansson O, Rutz S, Zetterberg H, Bauer E, Hähl T, Manuilova E, Mert MC, Wahl S, Blennow K, Stomrud E
Purpose
Conclusion
Recovery of amyloid β (Aβ)1-42 from CSF was approximately 100%, with little variability between replicates when 2.5 or 1.66 mL CSF was placed in Low-Bind Sarstedt False Bottom (FBT) tubes but was higher or lower than expected or more variable when other tubes or volumes were used. Generally, levels of Aβ1-42 were comparable between aliquots collected by gravity drip and lumbar fluid manometer (LFM) and over the course of lumbar puncture but the first collection by LFM was more variable. While effects of mixing, storage, and transport were observed, they could be attenuated by tube filling, storage/transport at 2-8°C, and/or transporting tubes upright. Levels of Aβ1-42 were largely unaffected by contamination with blood or centrifugation. Overall, similar results were observed for Aβ1-40 as Aβ1-42 but the magnitude of the changes were smaller and occasionally in the opposite direction. Phospho (P)-tau and Total (T)-tau levels were less affected by preanalytical factors. The authors conclude that CSF should be collected as >2.5 mL aliquots in Low-Bind FBT tubes and transported/stored at 2-8°C.
Studies (2)
Purpose
This study investigated the effect of tube type and fill volume on the measurement of biomarkers in CSF. Commercially obtained CSF was thawed for 30 min and 2.5, 1.66, and 0.75 mL were aliquoted into Sarstedt Low-Bind FBT tubes; 1.23 and 0.65 mL in Sarstedt 2.0 mL tubes; 2.0, 1.61, and 1.0 mL in Nunc tubes; 1.0, 0.84, and 0.5 mL in Eppendorf tubes (LoBind and Snap-Cap); and 0.63, 0.5, and 0.35 mL in Sarstedt 0.5 mL and Nalgene 0.5 mL tubes. Levels of Aβ1-42, Aβ1-40, P-tau, and T-tau were measured using Elecsys immunoassays on a Cobas e601 Analyzer.
Summary of Findings
Recovery of Aβ1-42 from CSF was approximately 100% with little variability between replicates when CSF was in Low-Bind Sarstedt FBT tubes filled with 2.5 or 1.66 mL but higher than expected when Starstedt 2.0 mL tubes were filled with 1.23 mL and when Nunc tubes were filled with 1.61 mL and lower than expected when Low-Bind Sarstedt FBT tubes were filled with 0.75 mL, Ependorf SnapCap tubes were filled with 0.5 or 1.0 mL, Sarstedt 2.0 tubes were filled with 0.65 mL, Nunc tubes were filled with 2.0 mL, or when Sarstedt or Nalgene 0.5 mL tubes were used, regardless of volume. Recovery of Aβ1-42 from CSF was 95-105% for all replicates, regardless of volume, when Eppendorf LoBind tubes were used but variability was higher than that in the Sarstedt Low-Bind FBT tubes. Importantly, recovery was lower in non-Low-Bind Sarstedt FBT tubes than in Sarstedt Low-Bind tubes, though the magnitude of the effect was small. There was no effect of sterilization of Sarstedt FBT tubes on recovery of Aβ1-42 from CSF.
Biospecimens
FluidCerebrospinal FluidPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
ProteinClinical chemistry/auto analyzerPeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution Eppendorf LoBind tubeEppendorf Snap-Cap tubeLow-Bind Sarstedt FBT tubeNalgene 0.5 mL tubeNunc tubeStarstedt 0.5 mL tubeStarstedt 2.0 mL tubeStarstedt FBT tubeBiospecimen Aliquots and Components Aliquot size/volume 0.35 mL0.5 mL0.63 mL0.65 mL0.84 mL1.0 mL1.23 mL1.6 mL2.0 mL2.5 mL
Purpose
This study investigated the effects of collection method, gradient, mixing, centrifugation, storage, and transportation position on the measurement of Aβ1-42, Aβ1-40, P-tau, and T-tau in CSF. CSF was obtained from patients undergoing lumbar puncture for suspected normal pressure hydrocephalus. CSF was collected using a LFM and analyzed within 4-6 h of collection unless otherwise specified. To assess the effect of collection method, eight aliquots of CSF were collected from each of four patients (the first four by gravity drip and the last four by lumbar fluid manometer) and levels were compared to the mean of the four gravity drip aliquots. Freshly collected CSF Lumber puncture gradient effects were investigated by comparing levels in aliquots obtained over the course of obtaining 500 mL. The effect of mixing was investigated by filling tubes with 0.75 mL and 1.75 mL of freshly collected CSF from four patients and not mixing, mixing by roller for 15 min, and mixing by vortex. Stability was tested by storing CSF from six patients at room temperature for 25 h, 49 h, 6 days, or 8 days and at 2-8°C for 25 h, 49 h, 8 days, 14 days, or 15 days and comparing levels to those measured 4-6 h after collection. The effect of blood contamination was studied by adding 0, 0.1%, or 1% of patient blood stored at room temperature for 0 or 6 days and at 2-8°C for 14 days and compared to uncontaminated, fresh specimens. The effect of centrifugation was investigated by comparing Aβ1-42 concentrations by splitting the first 2 mL of CSF collected and one part centrifuged (conditions not specified) and the other not. The effect of transportation was investigated by transporting 1.5 mL and 3 mL aliquots of CSF at room temperature and 2-8°C upright or horizontally for 0, 24, or 40 h on a roller or inverter for 40 h. Levels of Aβ1-42, Aβ1-40, P-tau, and T-tau were measured using Elecsys immunoassays on a Cobas e601 Analyzer.
Summary of Findings
Although the first aliquot obtained by LFM in each of the four patients had lower levels of Aβ1-42 than the average of the aliquots collected by gravity drip, the subsequent aliquots obtained by LFM and those obtained by the drip method all had comparable levels of Aβ1-42. Levels of Aβ1-42 were comparable over the course of lumbar puncture. Mean levels of Aβ1-42 were higher when not mixed than when mixed, but the effect was fill volume-dependent when mixed with less of an effect noted when tubes containing 1.75 mL were vortexed than when the tubes contained 0.75 mL. Levels of Aβ1-42 remained within 5% of baseline when CSF was stored for up to 49 h at RT or 15 days at 2-8°C but longer storage resulted in increased variability in levels. Levels of Aβ1-42 were unaffected by storage upright for 40 h but declined in 1.5 mL aliquots when transported horizontally or stored on a roller or inverter, though the effects were larger in 1.5 mL aliquots than 3.0 mL aliquots and when transported at room temperature rather than 2-8°C such that no reduction was observed when 3.0 mL aliquots were transported or stored at 2-8°C under any of the tested conditions. Further investigation showed no effect of transportation position for aliquots of 2.25-3.0 but more variability in Aβ1-42 levels when aliquots of 2.25-3.0 mL were stored at room temperature than 2-8°C. Levels of Aβ1-42 remained within 10% of the fresh uncontaminated specimen when specimens were contaminated with up to 1% blood. Similarly, Aβ1-42 levels were very strongly correlated between matched aliquots that were centrifuged and not centrifuged (Spearman ρ=0.979). Overall, similar results were observed for Aβ1-40 as Aβ1-42 but the magnitude of the changes were smaller and occasionally in the opposite direction. P-tau and T-tau levels were less affected by preanalytical factors. The authors conclude that CSF should be collected in >2.5 mL aliquots in Low-Bind FBT tubes and transported/stored at 2-8°C.
Biospecimens
FluidCerebrospinal FluidPreservation Types
None (Fresh)Diagnoses
Other diagnosesPlatforms
ProteinClinical chemistry/auto analyzerPeptideClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Method of fluid acquisition Lumbar fluid manometerLumbar punctureBiospecimen Aliquots and Components Aliquot sequential collection Serial collections comparedBiospecimen Aliquots and Components Aliquot size/volume 0.75 mL1.5 mL1.75 mL3.0 mLBiospecimen Aliquots and Components Biospecimen components 0% blood added0.1% blood added1% blood addedBiospecimen Aliquots and Components Biospecimen mixing InverterRollerVortexBiospecimen Aliquots and Components Centrifugation CentrifugedNot centrifugedStorage Storage conditions HorizontallyUprightStorage Storage duration 14 days15 days25 h4-6 h49 h6 days8 daysStorage Storage temperature 2-8°CRoom temperatureClinical chemistry/auto analyzer Specific Targeted peptide/protein Aβ1-40Aβ1-42P-tauT-tau