Comparison of Three Blood Collection Tubes for 35 Biochemical Analytes: The Becton Dickinson Barricor Tube, Serum Separating Tube, and Plasma Separating Tube.

Shin S, Oh J, Park HD

Publication: Ann Lab Med (2021) • Vol 41 • Page 114-119PMID: 32829587View on PubMed

Purpose

This paper compared the levels of 35 analytes in blood collected from healthy patients in Barricor tubes to matched blood collected in Vacutainer Serum Separating Tubes II (SSTII) and plasma Separating Tubes (PST).

Conclusion

A lower percentage of specimens in Barricor tubes required manual removal of gel globules or fibrin after centrifugation than SST specimens and the average turn-around time was faster for Barricor tubes than either SST or PST. The desired bias limit was exceeded in specimens in Barricor tubes for lactate dehydrogenase (LDH), albumin, aspartate aminotransferase (AST), calcium, glucose, potassium, phosphorus, sodium, total protein, folate, and free thyroxine when compared to the SST specimen but only AST, calcium, carbon dioxide (CO2), and LDH exceeded the desired bias limit when compared to the PST specimen. Importantly, the bias in albumin, iron, LDH, phosphorous, and ferritin in Barricor tubes exceeded the desirable bias at one or more medical decision points but only the bias in LDH was greater than the desirable bias at all medical decision points compared to both the SST and PST specimens.

Studies (1)

Purpose

This study compared the levels of 35 analytes in blood collected from healthy patients in Barricor tubes to matched blood collected in Vacutainer Serum Separating Tubes II (SSTII) and Plasma Separating Tubes (PST). Blood from 30 healthy patients was drawn in random order into Barricor, Vacutainer SST II, and Vacutainer PST tubes. Blood in SSTs was allowed to clot for 30 minutes before centrifugation at 2,000 × g for 10 minutes at 25°C. Blood in Barricor tubes and PSTs was centrifuged immediately after blood collection at 1,300 × g for 10 minutes at 25°C. Blood in PSTs was centrifuged immediately after blood collection 3,500 × g for 5 minutes at 25°C. Levels of 27 clinical chemistry analytes were measured using a Roche Cobas c702 chemistry analyzer. Levels of cardiac tropo­nin I (cTnI) and creatine kinase-myocardial band (CK-MB) were measured with an ADVIA Centaur XP Immunoassay System. Turnaround time was recorded for each step in the process. Results for the specimens in SSTII and PST were compared to Barricor tubes. Percent bias was calculated using a Passing-Bablok slope of zero or one outside the 95% confidence interval and compared to the desired bias from the Westgard database. Clinical sig­nificance was based on the lower of the desirable biological variation database or total allowable error according to the Clinical Labo­ratory Improvement Amendments (CLIA) criteria.

Summary of Findings

A lower percentage of specimens in Barricor tubes required manual removal of gel globules or fibrin after centrifugation than SST tubes. The average turnaround time was faster for Barricor tubes than either SST or PST (8.8 min versus 43.6 min and 13.8 min, respectively) which the authors attribute to no coagulation time and shorter centrifugation. The desired bias limit was exceeded in specimens in Barricor tubes for LDH (-10.29), albumin (-1.9%), aspartate aminotransferase (-6.75%), calcium (-1.46%), glucose (4.78%), potassium (-4.41%), phos­phorus (-5.69%), sodium (-0.21%), total protein (1.94%), folate (45.13%), and free thyroxine (5.31%) when compared to the SST specimen, but only AST (8.82%), calcium (-1.16%), CO2 (-3.23%), and LDH (-9.86%) exceeded the desired bias limit when compared to the PST specimen. Importantly, the bias in albumin, iron, LDH, phosphorous, and ferritin in Barricor tubes exceeded the desirable bias at one or more medical decision points but only the bias in LDH was greater than the desirable bias at all medical decision points compared to both the SST and PST specimens.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    GasClinical chemistry/auto analyzer
    GlycoproteinImmunoassay
    ProteinClinical chemistry/auto analyzer
    ProteinImmunoassay
    PeptideClinical chemistry/auto analyzer
    PeptideImmunoassay
    Small moleculeClinical chemistry/auto analyzer
    Small moleculeImmunoassay
    LipidClinical chemistry/auto analyzer
    SteroidClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    Barricor tube
    Vacutainer PST Tube
    Vacutainer SST II