This paper investigated the effects of pneumatic tube system (PTS) transport on 19 markers of hemostasis in blood by comparing results of routine coagulation and coagulation factor activity tests, fibrinolytic parameters, and levels of anticoagulation indicators in matched specimens transported 500 m and 1000 m by PTS to those transported manually.
Falsely decreased FVIII activity following pneumatic tube transport.
Wang H, Wang L, Liang H, Wei J, Wu Y, Wang X, Xu J
Purpose
Conclusion
Although none of the specimens displayed hemolysis when transported manually and slight hemolysis was only observed in one of the thirty specimens transported halfway by PTS, 40% of the specimens transported 1000 m by PTS displayed hemolysis. Significant differences in the prothrombin time (PT) international normalized ratio (INR) were found in specimens transported by PTS 1000 m compared to matched manually transported specimens but the differences were not clinically significant and no significant effect of transport 500 m was observed. Levels of Factor (F)II, FV, FVII, FX, D-dimer, and Protein C were higher and FVIII, FIX, FXI, FXII, and α2-plasmin inhibitor (α2-PI) lower when specimens were transported by PTS rather than manually transported, regardless of distance of PTS transport. However, all biases were less than the clinical threshold with the exception of FVIII. The bias in FVIII was greater than the clinical threshold of 8.3% in 20 and 29 of the 30 specimens transported by PTS 500 m and 1000 m, respectively, but FVIII bias did not differ between hemolyzed and non-hemolyzed specimens. Significant differences between specimens transported by PTS 500 m versus 1000 m were also found for FII, FVII, FVIII, and FX. Activated partial thromboplastin time (APTT), fibrinogen (Fib), thrombin time (TT), FXI, fibrin/fibrinogen degradation products, plasminogen, and antithrombin did not differ among the groups.
Studies (1)
Purpose
This study investigated the effects of PTS transport on 19 markers of hemostasis in blood by comparing results of routine coagulation and coagulation factor activity tests, fibrinolytic parameters, and levels of anticoagulation indicators in matched specimens transported 500 m and 1000 m by PTS to those transported manually. Blood was collected in triplicate from 30 healthy volunteers (14 males and 16 females, aged 21-49 years) into Vacutainer tubes containing 3.2% sodium citrate. One tube from each volunteer was transported 1000 m manually to the laboratory, another tube was transported halfway by PTS (500 m) and manually for the remaining distance, and the final tube was transported the entire distance by PTS. PTS transport of specimens was in carriers with sponge rubber inserts through a system with a maximum speed of 7 m/sec. Specimens were evaluated for clot, volume, lipemia, and icterus before and after plasma isolation. Platelet-poor plasma was obtained by double centrifugation at 1500 x g for 15 minutes at 23°C. Hemoglobin levels were measured by spectrophotometer. PT, INR, APTT, and TT, activity of FII, FV, FVII, FVIII, FIX, FX, FXI, and levels of D-dimer, fibrin/fibrinogen degradation products, plasminogen and α2-PI, antithrombin and Protein C were measured using a Sysmex CS 5100 coagulation analyzer. All three specimens from each volunteer were processed and analyzed simultaneously. Clinical significance was predetermined based on the Clinical Laboratory Improvement Amendment of 1988 as 5.0% for PT and APTT, 6.7% for INR and fibrinogen, 8.3% for all clotting factor activities, and 10.0% for the other assays.
Summary of Findings
Although none of the specimens displayed hemolysis when transported manually and slight hemolysis was only observed in one of the thirty specimens transported halfway by PTS (4.4 mg hemoglobin/dL), 40% of the specimens (12 of 30) transported 1000 m by PTS displayed hemolysis with an average hemoglobin level of 8.4 mg/dL. PT was slightly shorter and the resultant INR lower in specimens transported by PTS 1000 m than in manually transported specimens (P=0.003 and P=0.022, respectively) but the differences were not clinically significant nor was there a difference in specimens transported 500 m by PTS. Levels of FII, FV, FVII, FX, D-dimer, and Protein C were higher and FVIII, FIX, FXI, FXII, and α2-PI lower when specimens were transported by PTS rather than manually transported, regardless of distance of PTS transport (P<0.05, all). However, all biases were less than the clinical thresholds with the exception of FVIII. The bias in FVIII was greater than the predetermined clinical threshold of 8.3% in 20 and 29 of the 30 specimens transported by PTS 500 m and 1000 m, respectively. Importantly, bias in FVIII did not differ between hemolyzed and non-hemolyzed specimens. Significant differences between specimens transported by PTS 500 m versus 1000 m were also found for FII (P=0.003), FVII (P<0.001), FVIII (P<0.001), and FX (P<0.001) but APTT, Fib, TT, FXI, fibrin/fibrinogen degradation products, plasminogen. and antithrombin did not differ among the groups.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
NormalPlatforms
ProteinSpectrophotometryProteinHematology/ auto analyzerMorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Specimen transport duration/condition 0 m in PTS1000 m in PTS500 m in PTSStorage Within hospital transportation method Hand-deliveredPneumatic tube system