The Immunohistochemical Expression of Programmed Death Ligand 1 (PD-L1) Is Affected by Sample Overfixation.

Barberà A, Marginet Flinch R, Martin M, Mate JL, Oriol A, Martínez-Soler F, Santalucia T, Fernández PL

Publication: Appl Immunohistochem Mol Morphol (2020)PMID: 32134754View on PubMed

Purpose

This paper investigated the effects of fixation duration (12-96 h) on programmed death receptor ligand (PD-L1) staining using two different antibodies on tonsil tissue specimens.

Conclusion

There was no effect of fixation time up to 72 h on the number of sections determined to be suboptimally stained for PD-L1 but a significantly higher percentage of specimens fixed for 96 hours were classified as suboptimal for both PD-L1  antibodies.

Studies (1)

Purpose

This study investigated the effects of fixation duration on programmed death receptor ligand (PD-L1) staining using two different antibodies on tonsil tissue specimens. Twenty-two 0.5 cm thick specimens obtained during tonsillectomy (no diagnosis provided) were fixed in neutral buffered formalin for 12, 24, 48, 72, or 96 h (including the final 120 minutes of in-processor fixation) and paraffin-embedded in an automated processor. Three 3 mm tissue cores from each specimen were arrayed into a paraffin block and sectioned. Sections were stained in an autostainer with the SP142 and SP263 antibodies against PD-L1. Two pathologists evaluated each spot and classified specimens as optimally stained when 90% of the observed macrophages and T-cells had moderate to intense staining (+1 or +2) in germinal centers, there was intense staining (+2) of some epithelial squamous cells, and no staining in cells of interfollicular regions with no background staining. Sections not meeting these criteria were determined to be suboptimally stained.

Summary of Findings

A significantly higher percentage of specimens fixed for 96 hours were classified as suboptimally stained compared to specimens fixed for 12-72 h h (15.6% versus <6%, P<0.0001). When data was separated by antibody, 20.8% and 10.4% were suboptimally stained when fixed for 96 h using the SP142 antibody and SP263 antibodies, respectively (P=0.046). There were no statistical differences between the two antibodies in number of sections determined to be suboptimally stained for fixation times from 12-72 h (range from 2.8%-6.1%).

  • Biospecimens

    TissueTonsil
  • Preservation Types

    Formalin
  • Diagnoses

    Not specified
  • Platforms

    ProteinImmunohistochemistry
    ProteinTissue microarray
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationTime in fixative
    12 h
    24 h
    48 h
    72 h
    96 h
    Immunohistochemistry SpecificTargeted peptide/protein
    PD-L1
    Immunohistochemistry SpecificType of antibody
    Two antibodies compared