Impact of preanalytical storage on the measurement of erythrocyte sedimentation rate using an infrared microphotometer system (TEST1).

Ilardo C, Richerd C, Rostain V

Publication: Scand J Clin Lab Invest (2020) • Page 1-2PMID: 32643450View on PubMed

Purpose

The purpose of this paper was to investigate the effects of storage temperature and duration on the erythrocyte sedimentation rate (ESR) using an infrared microphotometer system.

Conclusion

The mean ESR was significantly lower after 24 h storage at room temperature compared to storage <2 h at room temperature but was not significantly reduced after 8 h at room temperature or 24 h versus <2 h at 2-8°C.

Studies (1)

Purpose

The purpose of this study was to investigate the effects of storage temperature and duration on the ESR using an infrared microphotometer system. EDTA blood was collected from 102 patients (diagnosis unspecified) and stored at room temperature (15-25°C) and analyzed after <2 h, 8h, and 24 h. EDTA blood from 112 patients (diagnosis unspecified) was stored at 2-8°C and analyzed after <2 h, 8h, and 24 h. ESR was determined by optical density on an infrared microphotometer system.

Summary of Findings

The mean ESR was significantly lower after 24 h storage at room temperature compared to storage <2 h at room temperature (3.0 versus 5.5, P=0.019). Only non-significant declines in the mean ESR were noted after storage for 8 h compared to <2 h at room temperature (4.0 versus 5.5, P=0.512) or 24 h compared to <2 h at 2-8°C (7.0 versus 8.0, P=0.280).

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    MorphologySpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageStorage duration
    24 h
    8 h
    <2 h
    StorageStorage temperature
    15-25°C (Room temperature)
    2-8°C