The purpose of this paper was to compare activated partial thromboplastin time (APTT) and platelet counts in specimens obtained by centrifugation at 2000 x g versus 3000 x g.
Influence of sample centrifugation on plasma platelet count and activated partial thromboplastin time using patient samples.
Jesting A, Kemp Jacobsen K, De Cock A, De Preester H, Overgaard Foss Jensen K, Frank Joergensen S
Purpose
Conclusion
As expected, plasma platelet count was significantly lower when specimens were centrifuged at 3000 x g versus 2000 x g resulting in a much higher percentage of specimens meeting the criteria for platelet poor plasma (PPP). APTT was comparable between specimens centrifuged at 2000 x g and 3000 x g for the 49 patients with APTT <37 seconds. Although a statistically significant effect of centrifugation speed was observed in the 21 specimens with prolonged APTT (>37 seconds), there was no clear trend in the effect and the differences remained within the maximal acceptable bias of 10%. Additionally, APTT was very strongly correlated and showed good agreement between specimens centrifuged at 2000 x g and 3000 x g.
Studies (1)
Purpose
The purpose of this study was to compare APTT and platelet counts in specimens obtained by centrifugation at 2000 x g versus 3000 x g. Blood was collected from 70 patients referred for activated partial thromboplastin time (APTT) testing into two 3.2% citrate Vacuette tubes. Within 2 h of blood collection, one tube was centrifuged at 2000 x g for 10 min while the other was centrifuged at 3000 x g for 10 min. APTT was measured immediately using the APTT-SP Kit on an ACL TOP 750 analyzer. The top part of the plasma was removed and platelet counts were analyzed using a Sysmex XN 9000.
Summary of Findings
As expected, plasma platelet count was significantly lower when specimens were centrifuged at 3000 x g versus 2000 x g (P<0.001). Importantly, while 64% of specimens centrifuged at 3000 x g had a platelet count of <10 x 109 /L (the threshold for platelet poor plasma), only 6% of specimens centrifuged at 2000 x g met this threshold. APTT was comparable between specimens centrifuged at 2000 x g and 3000 x g for the 49 patients with APTT <37 seconds. While there was no clear effect of centrifugation speed on APTT for the 21 patients with prolonged APTT (>37 seconds), the APTT was significantly different between these specimens centrifuged at 2000 x g versus 3000 x g (P=0.025). Importantly, APTT was very strongly correlated (r=0.997) and showed good agreement between specimens centrifuged at 2000 x g and 3000 x g. All differences in APTT remained within the maximal acceptable bias of 10%.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
Cell count/volumeHematology/ auto analyzerMorphologyHematology/ auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Aliquots and Components Centrifugation Multiple speeds compared