The purpose of this paper was to investigate the stability of SARS-CoV-2, the virus that causes the novel 2019 coronavirus disease (COVID, COVID-19, COVID19) viral RNA in different media over seven days of frozen or refrigerated storage.
Evaluation of saline, phosphate buffered saline and minimum essential medium as potential alternatives to viral transport media for SARS-CoV-2 testing.
Rodino KG, Espy MJ, Buckwalter SP, Walchak RC, Germer JJ, Fernholz E, Boerger A, Schuetz AN, Yao JD, Binnicker MJ
Purpose
Conclusion
Cycle threshold (CT) values were within 2 cycles for all the specimens regardless of storage media and duration, indicating no change in sensitivity or stability over the 7-day period at either temperature. One specimen (stored in PBS) had lower CT values on days 1, 3, and 7 which the authors state may be due to variation in sample preparation. The CT values differed slightly between the two methods but both SARS-CoV-2 amplicons were detected in all specimens using both assays.
Studies (1)
Purpose
The purpose of this study was to investigate the stability of SARS-CoV-2 in different media over seven days of frozen or refrigerated storage. A total of 48 nasopharyngeal swabs from patients who tested negative for SARS-CoV-2 were placed into tubes containing 3 mL of M4-RT viral transport media (VTM), minimum essential media (MEM), saline, or PBS. Forty specimens (10 in each media type) were subsequently spiked with SARS-CoV-2 positive material at a concentration of 10,000 copies/mL, while the remaining eight specimens were kept as negative controls. Tubes were stored at 2 to 8°C or -15 to -25°C for 0, 1, 3, or 7 days before testing. SARS-CoV-2 was detected using two different real-time PCR-based assays: the Roche Cobas amplification of the open reading frame and envelope RNA and LDT SARS-CoV-2 amplification of the open reading frame and nucleocapsid RNA. All results were compared to those of the specimen in VTM.
Summary of Findings
CT values were within 2 cycles for all the specimens regardless of storage media and duration, indicating no change in sensitivity or stability over the 7-day period at either temperature. One specimen (stored in PBS) had lower CT values on days 1, 3, and 7 which the authors state may be due to variation in sample preparation. The CT values differed slightly between the two methods but both SARS-CoV-2 amplicons were detected in all specimens using both assays.
Biospecimens
CellNasopharynxPreservation Types
None (Fresh)FrozenDiagnoses
Pneumonia/Respiratory InfectionPlatforms
RNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Short-term storage solution M4-RT VTMMEMPBSSalineStorage Storage duration 0 days1 day3 days7 daysStorage Storage temperature -15 to -25°C2-8°CReal-time qRT-PCR Specific Targeted nucleic acid SARS-CoV-2 envelopeSARS-CoV-2 nucleocapsid, EnvelopeSARS-CoV-2 open reading frameReal-time qRT-PCR Specific Technology platform LDT SARS-CoV-2 amplificationRoche cobas