The purpose of this paper was to compare routine urinalysis and urinary sediment analysis in matched preserved and unpreserved first-morning urine.
Impact of chemical preservative in urine samples.
Salazar-García S, Lares-Villaseñor E, Bárcenas-Morales A, Vargas-Morales Juan M
Purpose
Conclusion
Specimens in the preservative tube had significantly higher pH and specific gravity than those without preservative. However, there was no difference in glucose, bilirubin, ketones, erythrocytes, proteins, nitrites, or leukocytes between specimens with and without chemical preservation. Further, microscopic analysis showed comparable numbers of leukocytes, erythrocytes, epithelial cells, mucin filaments, crystals, and bacteria in each specimen type.
Studies (1)
Purpose
The purpose of this study was to compare routine urinalysis and urinary sediment analysis in matched preserved and unpreserved first-morning urine. First morning urine was collected by 50 patients (diagnosis not specified) into BD Vacutainer Urinalysis Cups. After collection, the patient transferred urine to tubes containing sodium propionate 94%, ethyl paraben 5.6%, and chlorhexidine 0.4% that were included in the BD Vacutainer Urinalysis Kit. The tube and the remaining urine in the cup were delivered to the laboratory. Levels of glucose, bilirubin, ketones, specific gravity, erythrocytes, pH, proteins, nitrites, and leukocytes were measured using Multistix 10 SG test strips and a CLINITEK Advantus autoanalyzer 90 minutes after sampling. Analytical performance was validated using SIEMENS CheK-Stix Combo Pak Control Strips for Urinalysis. Urine was sedimented by centrifugation at 400 x g for 5 min, and the resultant pellet was resuspended and placed on a slide. The leukocytes, erythrocytes, epithelial cells, mucin filaments, crystals, and bacteria were evaluated using a microscope at 100X and 400X. A cut-off of P<0.05 was used in the Wilcoxon signed-rank test to determine the significance of differences between specimens with and without preservatives.
Summary of Findings
Specimens in the preservative tube had higher pH and specific gravity than those without preservative (P=0.001, both). Fewer specimens had a pH of 5 when chemically preserved than when not preserved (1 versus 17). However, there was no difference in glucose, bilirubin, ketones, erythrocytes, proteins, nitrites, or leukocytes between specimens with and without chemical preservation. Further, microscopic analysis showed comparable numbers of leukocytes, erythrocytes, epithelial cells, mucin filaments, crystals, and bacteria in each specimen type.
Biospecimens
FluidUrinePreservation Types
None (Fresh)Other PreservativeDiagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerProteinClinical chemistry/auto analyzerCell count/volumeClinical chemistry/auto analyzerSmall moleculeClinical chemistry/auto analyzerCell count/volumeLight microscopyMorphologyLight microscopyPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation ChlorhexadineEthylparabenNone (fresh)Sodium propionate