Quality of blood samples collected at home does not affect clinical decision making for the administration of systemic cancer treatment.

Cool L, Callewaert N, Pottel H, Mols R, Lefebvre T, Tack L, Lycke M, Missiaen J, Debruyne P, Vandijck D, Van Eygen K

Publication: Scand J Clin Lab Invest (2020) • Vol 80 • Page 215-221PMID: 32282290View on PubMed

Purpose

The purpose of this paper was to compare hematological and clinical chemistry analytes in specimens tested immediately after collection to those subjected to transport and delayed processing in a simulation of home collection.

Conclusion

Significant changes in the slopes for the hematocrit (HCT), platelet (PLT) count, and C-reactive protein (CRP) and the intercept for HCT, mean corpuscular volume (MCV), and CRP were found indicating proportional and constant deviation as a result of delayed analysis. The intra-individual variability in HCT and MCV but not PLT and CRP exceeded the cut-off for desirable analytical bias, but none of the changes affected whether patient reached the cut-off values for clinical decision making. Red blood cell (RBC), hemoglobin (Hb), white blood cell (WBC) count, absolute neutrophil count (ANC), creatinine, estimated glomerular filtration rate (eGFR), alanine aminotransferase (ALT), lactate dehydrogenase (LDH), and aspartate aminotransferase (AST) were comparable between matched specimens sent to the laboratory by pneumatic tube and analyzed immediately and those subjected to a delay and transported by car.

Studies (1)

Purpose

The purpose of this study was to compare hematological and clinical chemistry analytes in specimens analyzed immediately after collection to those subjected to transport and delayed processing in a simulation of home collection. Blood specimens were collected from 49 patients with cancer into K2EDTA vacutainer and serum tubes with gel separator between 8:00 and 10:30 AM. Matched specimens were transported to the laboratory using a pneumatic tube system (mean delay 19 min) or placed in a box at room temperature and transported around the hospital by car simulating home collection before arrival at the laboratory (mean delay 291 minutes). RBC count, hemoglobin, HCT, MCV, WBC count, ANC, and PLT counts were measured in EDTA blood using a XN 3000 cell counter. Serum was obtained by centrifugation (speed and duration not specified) and creatinine, AST, ALT, LDH, potassium, and CRP levels were measured using a Cobas 6000 instrument. The eGFR was calculated based on serum creatinine. The low and high hemolytic index cutoffs for LDH were 15 and 30 mg/dL Hb, respectively. Statistical significance was determined by the 95% confidence interval and visualized using Bland-Altman plots. The desirable bias was calculated from the European Federation of Clinical Chemistry and Laboratory Medicine (EFLM) database for biological variation.

Summary of Findings

The slopes for HCT, PLT count, and CRP levels in matched specimens were found to be significantly different from 1 (1.08, 1.04, and 1.02; respectively), indicating that the changes with delayed analysis were proportional to the amount of analyte present. Further, the intercept for HCT, MCV, and CRP were not 0 (-2.2, 6.71, and -0.10; respectively), indicating the deviation was constant. Importantly, the intra-individual variability in HCT and MCV exceeded the cut-off for desirable analytical bias (2.07% versus ±1.51% and 1.90% versus ±1.32%), but the intraindividual variability in PLT and CRP remained within the desirable analytical bias (2.54 versus ±5.07 and 2.54 versus 0.24 ±21.80, respectively). The change in HCT, MCV, PLT count, or CRP levels did not result in a change in whether the cut-off values for clinical decision making were reached for any of the patients. RBC, Hb, WBC, ANC, CREAT, eGFR, ALT, AST, and LDH were comparable between specimens sent to the laboratory by pneumatic tube and analyzed immediately and those subjected to a delay and transported by car. An outlier in the AST and LDH analysis was shown to have a high hemolysis index in both specimens of the pair (33 and 146 mg/dL Hb), but after exclusion of this specimen, there still was not a significant difference in AST or LDH between the groups. 

  • Biospecimens

    FluidBlood
    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    NeoplasticCarcinoma
    NeoplasticNot specified
  • Platforms

    Cell count/volumeHematology/ auto analyzer
    MorphologyHematology/ auto analyzer
    ProteinClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageBetween site transportation method
    Car/Truck
    Not transported
    StorageTime at room temperature
    Average of 19 min
    Average of 291 min