DNA extraction from FFPE tissue samples - a comparison of three procedures.

Sarnecka AK, Nawrat D, Piwowar M, Ligęza J, Swadźba J, Wójcik P

Publication: Contemp Oncol (Pozn) (2019) • Vol 23 • Page 52-58PMID: 31061638View on PubMed

Purpose

The purpose of this study was to investigate differences due to DNA extraction kit choice on the yield and purity of DNA from archival formalin-fixed paraffin-embedded (FFPE) blocks.

Conclusion

Average DNA concentrations and purity (A260/280 ratios) were significantly higher when extraction was with the Cobas or MAXwell kits rather than the QIAamp Kit. While concentrations were generally higher when extraction was with the Maxwell Kit than the Cobas Kit, the differences were not significant.

Studies (1)

Purpose

The purpose of this study was to investigate the effects of DNA extraction kit choice on the yield and purity of DNA from archival FFPE blocks. Forty-two archival FFPE blocks containing melanoma, colon cancer, or lung adenocarcinoma were included in the study but no processing or storage details were provided. Fresh 5 µm sections were cut from each block and a representative slide was H&E stained. Unstained sections were deparaffinized in xylene and tumor-rich areas were microdissected from the slide using the H&E section as a guide. DNA was extracted from equal amounts of tissue using the QIAamp DNA FFPE Tissue Kit, the Cobas DNA Sample Preparation Kit, or the Maxwell Kit according to the manufacturer’s directions. DNA was quantified spectrophotometrically using a NanoDrop and using the double-stranded Qubit 3.0 Kit. DNA purity was evaluated based on spectrophotometric ratios.

Summary of Findings

The average DNA concentration was significantly lower when extraction was with the QIAamp Kit than the Cobas Kit (using spectrophotometer 18.00 versus 50.50 ng/µL, P<0.0001 and using Qubit 4.79 versus 9.15 ng/µL, P=0.0009) or MAXwell Kit (using spectrophotometer 18.00 versus 102.72 ng/µL, P<0.0001 and using Qubit 4.79 versus 31.28 ng/µL, P=0.0120), but levels were modestly correlated between the QIAamp and other extraction methods (ρ=0.51-0.71). While concentrations were generally higher when extraction was with the Maxwell Kit than the Cobas Kit, the differences were not significant and levels were strongly correlated between methods when quantified by spectrophotometer (ρ=0.85) or Qubit (ρ=0.77). The mean ratio of absorbance at 260 to 280 was lower when extracted with the QIAamp Kit than the Cobas Kit (1.78 versus 1.84, P=0.0003) or the Maxwell Kit (1.78 versus 1.82, P=0.0033), but the median ratio was highest when extraction was with the QIAamp Kit (1.90 versus 1.77 with Cobas and 1.81 with Maxwell).

  • Biospecimens

    TissueColorectal
    TissueLung
    TissueSkin
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticMelanoma
    NeoplasticCarcinoma
  • Platforms

    DNAFluorometry
    DNASpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationAnalyte isolation method
    Cobas DNA Sample Preparation Kit
    Maxwell 16 FFPE Plus LEV DNA Purification Kit
    QIAamp DNA FFPE Tissue Kit
    Spectrophotometry SpecificTechnology platform
    NanoDrop
    Qubit 3.0