Mean ionized calcium concentrations were significantly lower for all blood volumes collected in Na-Hep and Li-Hep tubes (P<0.0001, all) compared to SST specimens, but only the Na-Hep tubes that were 87.5% (0.5 mL) and 75% (1 mL) underfilled and only the Li-Hep tubes that were 75% (0.5 mL) and 50% (1 mL) underfilled showed a clinically significant difference (≥0.1 mmol/L) compared to SST specimens. Importantly, the mean differences in ionized calcium concentration in Li-Hep tubes were significantly lower than Na-Hep tubes containing the same volume of blood (P<0.001). While the Na-Hep and Li-Hep tubes collected with 2.0 mL did not have a statistically significant difference in mean pH compared to SST specimens, significant differences were observed for the Na-Hep tubes containing 0.5 mL, 1.0 mL, and 4.0 mL of blood and Li-Hep tubes with 0.5 and 1 mL of blood (P<0.0001, all), but only the differences in Na-Hep and Li-Hep tubes filled with 0.5 mL of blood were clinically significant (≥0.1 pH unit from SSTs).