Impact of underfilling heparinized collection tubes on ionized calcium measurements.

Tung JK, Bowen RAR

Publication: Clin Chim Acta (2019)PMID: 31782994View on PubMed

Purpose

This paper investigated the effect of underfilling lithium heparin and sodium heparin blood collection tubes on the measurement of ionized calcium in blood. Ionized calcium levels in blood were compared to those in serum from properly filled specimens collected in serum separator tubes (SSTs).

 

Conclusion

Mean ionized calcium concentrations were significantly lower in lithium heparin (Li-Hep) tubes compared to sodium heparin (Na-Hep) tubes containing the same volume of blood.  Although mean ionized calcium concentrations were lower for all blood volumes in Na-Hep and Li-Hep tubes compared to SST specimens, the differences were only clinically significant in the the Na-Hep tubes that were 87.5% and 75% underfilled and only the Li-Hep tubes that were 75% and 50% underfilled. Significant differences in pH were observed for blood collected in Na-Hep and Li-Hep tubes compared to serum from SST specimens, but the differences were only clinically significant in Na-Hep and Li-Hep tubes filled with 0.5 mL of blood.

Studies (1)

Purpose

This study investigated the effect of underfilling lithium heparin and sodium heparin blood collection tubes on the measurement of ionized calcium in blood. Ionized calcium levels in blood were compared to those in serum from properly filled specimens collected in serum separator tubes (SSTs). Whole blood was collected from 20 healthy volunteers using a 21-gauge butterfly needle into blood collection tubes at different volumes. First, a “full-draw” of blood was collected in a 5 mL serum separator tube (SST) as a control and then 4.0, 2.0, 1.0, and 0.5 mL was collected into separate 4 mL Vacutainer sodium heparin (Na-Hep) blood tubes and 2.0, 1.0, and 0.5 of blood was collected in separate 2 mL Vacutainer lithium heparin (Li-Hep) tubes. All tubes were thoroughly mixed to distribute the heparin and minimize the formation of microclots. The serum was allowed to clot in the SST tubes for 60 min before centrifugation at 4000 rpm for 5 min. Ionized calcium concentrations and pH were measured using a blood gas analyzer within 15 min of collection for the Na-Hep and Li-Hep tubes and 70 min for the SST tubes.

Summary of Findings

Mean ionized calcium concentrations were significantly lower for all blood volumes collected in Na-Hep and Li-Hep tubes (P<0.0001, all) compared to SST specimens, but only the Na-Hep tubes that were 87.5% (0.5 mL) and 75% (1 mL) underfilled and only the Li-Hep tubes that were 75% (0.5 mL) and 50% (1 mL) underfilled showed a clinically significant difference (≥0.1 mmol/L) compared to SST specimens. Importantly, the mean differences in ionized calcium concentration in Li-Hep tubes were significantly lower than Na-Hep tubes containing the same volume of blood (P<0.001). While the Na-Hep and Li-Hep tubes collected with 2.0 mL did not have a statistically significant difference in mean pH compared to SST specimens, significant differences were observed for the Na-Hep tubes containing 0.5 mL, 1.0 mL, and 4.0 mL of blood and Li-Hep tubes with 0.5 and 1 mL of blood (P<0.0001, all), but only the differences in Na-Hep and Li-Hep tubes filled with 0.5 mL of blood were clinically significant (≥0.1 pH unit from SSTs).

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Lithium heparin
    Sodium heparin
    Biospecimen AcquisitionType of collection container/solution
    2 mL Vacutainer lithium heparin tube
    4 mL Vacutainer sodium heparin blood tube
    5 mL serum separator tube (SST)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    0.5 mL
    1 mL
    2 mL
    4 mL
    5 mL
    Biospecimen Aliquots and ComponentsBlood and blood products
    Serum