Skip to main content

Investigating pre-analytical requirements for serum and plasma based infrared spectro-diagnostic.

Lovergne L, Lovergne J, Bouzy P, Untereiner V, Offroy M, Garnotel R, Thiéfin G, Baker MJ, Sockalingum GD

Publication: J Biophotonics (2019) • Vol 12 • Page e201900177PMID: 31276294View on PubMed

Purpose

The purpose of this study was to investigate how the infrared spectrometry spectra of serum and plasma specimens are influenced by collection tube type, sample volume, frozen storage duration, dehydration method, and relative humidity during drying. Methods to computationally compensate for the effects of humidity were also investigated.

Conclusion

A total of 1.4% of serum and 0.7% of plasma specimens did not pass the spectra quality test and 3.2% of serum and 0.3% of plasma specimens had increased absorbance in the region of 3700 to 3320 cm-1, which is indicative of water being present. Spectra of plasma and serum specimens did not cluster based on tube volume (0.2 or 1.5 mL), specimen volume (50 or 100 µL), or storage duration, but results clustered by patient. However, when spectra of serum and plasma specimens were normalized results did not cluster by patient, although a shift in principle component 1 (PC1) based on the relative humidity during drying was observed. Further analysis showed that this spectral shift was attributable to water, but effects was also observed in vacuum-dried plasma and serum, indicating that rehydration occurred during spectral acquisition despite the presence of desiccant in the spectrometer. Specimens clustered by patient when  the PC1 component was removed from the normalized data set or extended multiplicative signal correction (EMSC) was applied using the spectra of saline; individually, these steps also reduced the mean standard deviation (SD) and their combination (removal of PC1 and EMSC correction) reduced variation  variation between replicates even further and produced better separation among patients.

Studies (1)

Purpose

The purpose of this study was to investigate how the infrared spectra of serum and plasma specimens are affected by collection tube type, sample volume, frozen storage duration, dehydration method, and relative humidity. Methods to computationally compensate for the effects of humidity were also investigated. Blood was collected from three healthy patients into serum tubes and lithium heparin tubes. Serum tubes were allowed to clot for 30 min before processing. Serum and plasma were obtained by centrifugation at 3,000 rpm for 15 min, and then 50 µL aliquots were placed in 0.2 mL and 1.5 mL tubes, and 100 µL aliquots were placed in 1.5 mL tubes. All aliquots were stored at -80°C for 9 months, with two aliquots of serum and plasma of each volume/tube size thawed monthly. Specimens were diluted in saline and aliquoted as ten 5 µL samples and allowed to dry at room temperature (21.4-25.9°C, 30-60% relative humidity) or under vacuum for 90 min.  The spectra of the specimens were analyzed using the high-throughput module (HTS-XT) coupled to a Fourier-transform infrared (FTIR)-spectrometer. After quality testing spectra were reduced to the range of 4000 to 800 cm-1 and normalized using a rubberband baseline correction before vector normalization and principle component analysis.

Summary of Findings

A total of 1.4% of serum and 0.7% of plasma specimens did not pass the spectra quality test and 3.2% of serum and 0.3% of plasma specimens had increased absorbance in the region of 3700 to 3320 cm-1, with peaks at 3350, 3510, 3484, 3450 and 3420 cm-1 that were not observed in the other specimens. The authors concluded that these peaks indicate the presence of water, which may have been introduced when the specimens were dried at room temperature with a relative humidity >50%. There was no evidence of spectra from plasma or serum clustering based on tube size (0.2 or 1.5 mL) or specimen volume (50 or 100 µL). PCA revealed that  plasma spectra clustered by patient, with specimens from the same patient co-clustering over the 9 month period, indicating no significant changes occurred during storage of plasma at -80°C. However, serum normalized spectra did not cluster by patient but showed a shift based on relative humidity during drying and further analysis revealed that this was due to a negative peak at 3470 cm-1 that was attributable to water. Interestingly, when specimens were vacuum-dried plasma and serum samples still clustered based on date of analysis indicating that the rehydration occurred during spectral acquisition despite the presence of desiccant in the spectrometer. As PC1 was affected by water, when the PC1 component was removed from the normalized data set and the specimens were reanalyzed with all remaining PCS specimens clustered by patient and no longer separated based upon relative humidity. Further, removal of PC1 reduced the mean standard deviation (SD) for both plasma and serum specimens. When an extended multiplicative signal correction (EMSC) was applied using the spectra of saline the SD was also reduced but did not fully correct for relative humidity. However, the combined removal of PC1 and EMSC correction further reduced variation between replicates and resulted in better separation among patients.

  • Biospecimens

    FluidPlasma
    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    ProteinSpectrophotometry
    Small moleculeSpectrophotometry
    PeptideSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsAliquot size/volume
    100 µL
    50 µL
    Biospecimen PreservationDehydration duration/condition
    At room temperature (varying humidity)
    Under Vacuum
    StorageStorage conditions
    0.2 mL tube
    1.5 mL tube
    StorageStorage duration
    1 month
    2 months
    3 months
    4 months
    5 months
    6 months
    7 months
    8 months
    9 months
    Spectrophotometry SpecificData handling
    EMSC corrected based on saline spectra
    Not corrected
    PC1 removed
    Saline normalized