Insufficient mixing of thawed serum samples leading to erroneous results - experience from a field study and use of a correction procedure.

Ekström U, Apelqvist J, Hansson E, Bodin T, Wegman DH, Abrahamson M, Jakobsson K

Publication: Scand J Clin Lab Invest (2019) • Page 1-7PMID: 31847598View on PubMed

Purpose

This paper investigated the effects of the method of mixing serum specimens after frozen storage and storage of gel and non-gel tubes on levels of creatinine, sodium, cystatin C, and albumin.

Conclusion

A gradient in creatinine, sodium, cystatin C, and albumin levels was observed when specimens were not mixed or were only vortexed for 2-3 seconds (Standard Vortex Method) and in sodium, creatinine, and cystatin C in one of five specimens vortexed for 10 sec, but no gradient was observed in specimens inverted 10 times with or without additional vortexing. While specimens collected without gel only displayed a gradient when not mixed, those collected with gel and incubated at room temperature or 50°C displayed a gradient in analyte levels in three of four specimens subjected to the Standard Vortex Method. Importantly, as all analytes were similarly affected, it was possible to apply a correction resulting in differences of <1% between the sampling locations.

Studies (2)

Purpose

This study investigated the effects of the method of mixing serum specimens after frozen storage on levels of creatinine, sodium, cystatin C, and albumin. Venous blood was obtained from five fasting healthy volunteers into six serum tubes without gel. One tube was immediately analyzed (control) and the remaining five were incubated for 1 h at room temperature before centrifugation for 10 min at 2000 x g to obtain serum which was refrigerated for 2 h and then stored at -20°C for 4 days followed by storage at -80°C for 1.5 days. One tube was thawed 4 h, inverted 10 times, vortexed for 10 sec, and centrifuged at 2200 x g for 10 min (Optimal Method). The second tube was thawed for 1–2 h at what, mixed by manual inverting 10 times, then centrifuged at 2200 x g for 10 min (Manual Mixing Method). The third tube was thawed for 1–2 h, mixed by vortexing for 10 sec, then centrifuged at 2200 x g for 10 min (Extended Vortex Method). The fourth tube was thawed for 1–2, mixed by vortexing until a swirl was observed (2-3 sec), then centrifuged at 2200 x g for 10 min (Standard Vortex Method). The final tube was thawed 1-2 h and then centrifuged at 2200 x g for 10 min (No Mixing Method). After the mixing procedure, aliquots were obtained from the top, middle, and bottom. Levels of sodium, creatinine, cystatin C, and albumin were measured on a Cobas 701 clinical chemistry instrument.

Summary of Findings

A gradient was observed for all four analytes in each of the five specimens that were not mixed and those that were vortexed for only 2-3 seconds (Standard Method) with higher levels of each analyte found at the bottom of the tube than the middle or top. Levels of analytes were also higher in the middle versus top in all specimens that were not mixed or that were subjected to standard vortexing except one standard vortex specimen that had the same level of creatinine and cystatin C in the middle and top and higher levels of sodium and albumin in the middle than top. One of the five specimens subjected to vortexing for 10 sec showed a similar gradient in sodium, creatinine, and Cystatin C but not albumin. No gradients were observed in the specimens processed using the Optimal or Manual Mixing methods.

  • Biospecimens

    FluidSerum
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationSample mixing
    Inverted 10 times (Manual)
    Inverted 10 times and vortexed for 10 sec (Optimal Mixing)
    Not mixed
    Vortexed 10 sec (Extended Vortexing)
    Vortexed until swirl (Standard Vortexing)