Quality assessment on the long-term cryopreservation and nucleic acids extraction processes implemented in the andalusian public biobank.

Ortega-Pinazo J, Díaz T, Martínez B, Jiménez A, Pinto-Medel MJ, Ferro P

Publication: Cell Tissue Bank (2019) • Vol 20 • Page 255-265PMID: 30903409View on PubMed

Purpose

This paper investigated the effects of long-term frozen storage of peripheral blood mononuclear cells (PBMCs) on viability and RNA yield and quality. The effect of storing DNA at -80˚C on DNA yield and integrity was also examined.

Conclusion

Although the viability of PBMCs was 40% lower in the specimens stored frozen for 6-10 years than in the fresh specimens, there was no effect of storage duration on RNA yields, purity (OD 260/280), or integrity and RT-PCR success was unaffected. DNA yield, DNA quality as determined by ratio of yield by spectrophotometry to PicoGreen, and spectrophotometric purity (OD 260/280) were comparable in fresh and frozen specimens. Electrophoresis showed a band of around 23.1 kb, indicating intact DNA, and PCR amplification was successful in all specimens.

Studies (2)

Purpose

This study investigated the effects of long-term frozen storage of peripheral blood mononuclear cells (PBMCs) on viability and RNA yield and quality. Blood was collected from 70 patients at the Neurology Service of the Regional University Hospital of Malaga, Spain between 2008 and 2012 into an EDTA tube. Six additional blood specimens were collected in 2018 and analyzed fresh. PBMCs were isolated using a gradient density with LymphoSep and resuspended in RPMI media with fetal bovine serum (FBS). Live cells were stained with Trypan blue and counted. PBMCs were then cryopreserved with DMSO and frozen at -80˚C for 24 h and then stored at -196˚C. Viability was determined by the Trypan blue exclusion method. Cell were resuspended in Trizol and stored at -80˚C for 48 h before RNA was extracted using the miRNeasy mini kit. RNA purity was assessed using a NanoDrop and integrity by electrophoresis. Integrity was assessed by RT-PCR amplification of a 429 bp fragment of HIST1H4A.

Summary of Findings

The viability of PBMCs was 40% lower in the specimens stored frozen for 6-10 years than in the fresh specimens (P<0.003 for all time-points), but there was no effect of storage duration. RNA yields from PBMCs were comparable in fresh specimens and those frozen for 6-10 years and there was no effect on RNA purity (OD 260/280). Electrophoresis showed clear ribosomal RNA bands. RT-PCR showed similar amplification in all specimens.

  • Biospecimens

    CellWhite Cells
  • Preservation Types

    FrozenNone (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    Cell count/volumeLight microscopy
    RNASpectrophotometry
    RNAElectrophoresis
    RNART-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Frozen
    None (fresh)
    StorageStorage duration
    0 years
    10 years
    6 years
    7 years
    8 years
    9 years