Comparison of Roche Cell-Free DNA collection Tubes to Streck Cell-Free DNA BCTs for sample stability using healthy volunteers.
Parackal S, Zou D, Day R, Black M, Guilford P
Purpose
Conclusion
There was a high degree of inter-individual variability in cfDNA levels. Although cfDNA concentration was not significantly altered when blood was stored for up to 14 days in Streck tubes before centrifugation, significant increases were noted when stored in Roche tubes for 7 days or more. Capillary electrophoresis showed obvious degradation of specimens stored in EDTA tubes for 14 days but only a small decrease in the cfDNA peak in the Streck and Roche tube specimens. The authors conclude that genomic DNA contamination is largely avoided through the use of Streck or Roche tubes.
Studies (1)
Purpose
This study investigated the effects of storing blood in Streck and Roche cell-free DNA blood collection tubes on cfDNA yield and genomic DNA contamination and compared the electropherograms to those from blood stored in K2EDTA tubes. Blood was collected from 20 healthy volunteers into five Streck cell-free DNA BCTs and five Roche cell-free DNA BCTs andfrom one volunteer into K3EDTA Vacutainer tubes. Tubes were stored upright in a temperature controlled room (22˚C) for 0 days (<2 h from collection), 4 days, 7 days, 10 days, or 14 days. Plasma was obtained by centrifugation at 200 rcf for 10 min followed by 1600 rcf for 10 min with break and acceleration set to zero. The plasma was then transferred to a new tube and spun at 1600 rcf. DNA was extracted from plasma using the QIAamp Circulating Nucleic Acids Kit using centrifugation rather than vacuum and elution in DNAse-free/RNAse-free water rather than buffer. DNA concentration was determined using the Qubit dsDNA High Sensitivity Kit. Fragment size profile was evaluated by capillary electrophoresis on a bioanalyzer using the High Sensitivity DNA Analysis Kit.
Summary of Findings
There was a high degree of inter-individual variability in cfDNA levels. Analysis of the data using the Kruskal-Wallis test showed no association with the changes in DNA concentration across days with tube type. The Wilcoxon rank-sum test showed no evidence that the distribution of concentration for the time-points was not centered around zero, but paired analysis identified significant increases in concentration between Day 0 and Days 7, 10, and 14 when stored in Roche tubes (P=0.01, P=0.046, and P=0.0016, respectively) but not Streck tubes (P>0.05). There were no significant differences using the Wilcoxon signed-rank test when specimens in Streck tubes at Day 0 were compared to specimens in Roche tubes at any time-point (P>0.05).
Capillary electrophoresis showed obvious degradation of specimens stored in EDTA tubes for 14 days. For specimens stored in Streck tubes for 14 days, the cfDNA peak decreased and there was a slight increase in degradation products, but the cfDNA and other peaks (including a small genomic DNA peak) remained largely similar. For whole blood stored in Roche tubes, the only clear peak was from cfDNA and the peak was slightly smaller and the baseline more jagged after 14 days of storage, which the authors attribute to potential ethanol contamination. The authors conclude that genomic DNA contamination is largely avoided through the use of Streck or Roche tubes.
Biospecimens
FluidBloodFluidPlasmaPreservation Types
Streck/BCTOther PreservativeNone (Fresh)Diagnoses
NormalPlatforms
DNAFluorometryDNAAutomated electrophoresis/BioanalyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Acquisition Type of collection container/solution K2EDTA tubeRoche cfDNA BCTStreck cfDNA BCTBiospecimen Aliquots and Components Centrifugation Centrifugation delays investigatedStorage Storage duration 0 days10 days14 days4 days7 days