This paper investigated the effects of heating serum specimens prior to extraction, extraction method used, and real-time PCR array type on the number of microRNAs (miRNAs, miRs) detected and the distribution of cycle threshold (CT) values.
Data on the effect of heat and other technical variables on the detection of microRNAs in human serum.
Camacho L, Porter-Gill P, Silva CS
Purpose
Conclusion
CT values tended to be lower for the TaqMan Low Density Human microRNA Array A than the Exiqon MicroRNA Ready-to-use PCR Human Panel I V4.M. Extraction method had little effect on the number of miRNAs detected or the distribution of the CT values when the RNA was not heated before extraction. However, heating serum prior to extraction resulted in the detection of fewer miRNAs using either array when extracted using the SeraMir Exosome RNA Purification Kit but had no effect when extraction was with the miRCURY RNA Isolation Kit.
Studies (1)
Purpose
The purpose of this study was to investigate the effects of heating serum specimens prior to extraction, extraction method used, and real-time PCR array type on the number of miRNAs detected and the distribution of CT values. Four commercially-obtained frozen human serum specimens were divided into two aliquots each. One aliquot was heated at 60˚C for 120 min and the other was not heated. Heated and not heated aliquots were divided and miRNA was extracted using both the miRCURY RNA Isolation Kit for Biofluids and the SeraMir Exosome RNA Purification Kit. miRNA was reverse-transcribed using the TaqMan Micro-RNA Reverse Transcription Kit and quantified by the real-time PCR-based TaqMan Low Density Human microRNA Array A or reverse-transcribed using the miRCURY LNA Universal cDNA Synthesis Kit II and quantified using the Exiqon MicroRNA Ready-to-use PCR Human Panel I V4.M Array.
Summary of Findings
CT values tended to be lower for the TaqMan Low Density Human microRNA Array A than the Exiqon MicroRNA Ready-to-use PCR Human Panel I V4.M. The authors attribute this difference to the inclusion of a preamplification step in the Applied Biosystems protocol. Extraction method had little effect on the number of miRNAs detected or the distribution of the CT values when the RNA was not heated before extraction. However, heating serum prior to extraction resulted in the detection of fewer miRNAs using either array when extracted using the SeraMir Exosome RNA Purification Kit but had no effect when extraction was with the miRCURY RNA Isolation Kit.
Biospecimens
FluidSerumPreservation Types
FrozenDiagnoses
Not specifiedPlatforms
RNALow density arrayRNAReal-time qRT-PCRPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage duration 0 h2 h at 60˚CAnalyte Extraction and Purification Analyte isolation method SeraMir Exosome RNA Purification KitmiRCURY RNA Isolation Kit for BiofluidsReal-time qRT-PCR Specific Technology platform Exiqon MicroRNA Ready-to-use PCR Human Panel I V4.M ArrayTaqMan Low Density Human microRNA Array A