Skip to main content

Evaluation of the efficacy of various reagents in improving microRNA extraction.

Kikuchi A, Naruse A, Sawamura T, Nonaka K

Publication: Ann Clin Biochem (2019) • Vol 56 • Page 375-380PMID: 30813744View on PubMed

Purpose

This paper investigated if microRNA (miRNA, miR) extraction efficiency using the HighPure miRNA isolation kit could be improved through the use of alternative binding enhancers.

Conclusion

The mean cycle threshold values for miR-21 were 1.91-2.83 cycles lower using the alternative binding solutions than when the Roche binding enhancer solution was used, indicating an increased efficiency of extraction. The largest decrease in mean cycle threshold values were obtained using polyethylene glycol (PEG) 1540 and PEG 2000 (2.42 and 2.83, respectively).

Studies (1)

Purpose

The purpose of this study was to determine if miRNA extraction efficiency using the HighPure miRNA isolation kit could be improved through the use of an alternative binding enhancer. Ten surgically-resected FFPE colon cancer specimens were obtained (additional details not provided). miRNA was extracted from xylene deparaffinized 5 µm sections using the HighPure miRNA isolation kit in conjunction with the optimal volume (experimentally determined) of different binding enhancer solutions. The tested binding solutions included 100 µL of the Roche binding enhancer solution (BE), 400 µL of 2-butanol, 400 µL of 2-propanol, 300 µL of acetone, 300 µL of acetonitrile, 400 µL of ethanol, 200 µL of PEG 1540, 200 µL of PEG 2000, 300 µL of tetraethylene glycol dimethyl ether (TDE), 300 µL of tetrahydrofuran (THF), and 500 µL of 1,4-dioxane. miR-21 levels following extraction were determined by real-time RT-PCR and compared to that using the Roche binding enhancer solution.

Summary of Findings

The mean cycle threshold values for miR-21 were 1.91-2.83 cycles lower using the alternative binding solutions than when the Roche binding enhancer solution was used, indicating an increased efficiency of extraction. The largest decrease in mean cycle threshold values were obtained using PEG 1540 and PEG 2000 (2.42 and 2.83, respectively). PEG was seen as particularly advantageous as the optimal volume was lower than for the other solutions tested.

  • Biospecimens

    TissueColorectal
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticCarcinoma
  • Platforms

    RNAReal-time qRT-PCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationAnalyte isolation method
    HighPure miRNA isolation kit with 1,4-dioxane
    HighPure miRNA isolation kit with 2-butanol
    HighPure miRNA isolation kit with 2-propanol
    HighPure miRNA isolation kit with PEG 1000
    HighPure miRNA isolation kit with PEG 1540
    HighPure miRNA isolation kit with PEG 2000
    HighPure miRNA isolation kit with PEG 600
    HighPure miRNA isolation kit with TDE
    HighPure miRNA isolation kit with THF
    HighPure miRNA isolation kit with candidate binding solution
    HighPure miRNA isolation kit with ethanol
    HighPure miRNA isolation kit with the Roche binding enhancer solution
    Real-time qRT-PCR SpecificTargeted nucleic acid
    miR-21