Effect of Preanalytic Variables on an Automated PTEN Immunohistochemistry Assay for Prostate Cancer.

Guedes LB, Morais CL, Fedor H, Hicks J, Gurel B, Melamed J, Lee P, Gopalan A, Knudsen BS, True LD, Scher HI, Fine SW, Trock BJ, De Marzo AM, Lotan TL

Publication: Arch Pathol Lab Med (2019) • Vol 143 • Page 338-348PMID: 30295067View on PubMed

Purpose

This paper investigated whether PTEN immunostaining in benign and tumor prostate specimens is adversely affected by cold ischemia time (0.5 - 48 h); fixative (10% neutral buffered formalin, NBF; Bouins, Hollande); the duration of formalin fixation (4-48 h), FFPE block storage (2 weeks - 23 y), unstained FFPE slide storage (<1 month, 5 y); institution-specific tissue processing protocols; or automated staining platform choice. Specimens were assembled into a tissue microarray prior to analysis. Immunohistochemical staining was either visually scored by a pathologist blind to the experimental assignments or quantified using open source software.

Conclusion

PTEN immunostaining results were not significantly affected by cold ischemia time, duration of formalin fixation, or automated staining platform choice. Significant differences were observed among different fixatives as the intensity of PTEN immunopositive staining was significantly stronger among prostate specimens fixed in 10% NBF compared to Bouins (P=0.03) or Hollande's fixative (P=0.004). While a probable effect of prolonged FFPE block storage was observed among one sample set (P=0.02), analysis of a second sample set failed to confirm the finding. Equivalent PTEN immunostaining was observed among prostate biopsy specimens that underwent tissue processing at 11 different institutions according to institution-specific protocols.

Studies (6)

Purpose

The purpose of this study was to determine the effects of cold ischemia time (the time between specimen excision and fixation) on intensity of PTEN immunohistochemical staining in formalin-fixed, paraffin-embedded benign prostate tissue specimens. Five biopsy cores (7 mm) were collected from 10 different prostates obtained by radical prostatectomy and subjected to 0, 1, 2, 4, or >4 hours of cold ischemia at room temperature in a humidified chamber prior to preservation. Tissue was preserved in 10% neutral buffered formalin (NBF) for 24 h, processed, and paraffin-embedded. A second sample set of biopsies collected from three radical prostatectomies containing benign tissue that were subjected to a cold ischemia time at room temperature of 0.5 to 48 h was also evaluated. Between two and four cores (0.6 mm) were punched from each FFPE block and assembled into a tissue microarray block. Automated immunohistochemical staining of PTEN was detected after antigen retrieval with Epitope Retrieval Solution 2 for 20 min using the Leica Bond platform. Immunostaining was quantified using FrIDA TMAJ software.

Summary of Findings

Cold ischemia time and the intensity of PTEN immunostaining in benign prostate FFPE specimens were not correlated in the first sample set of specimens subjected to a cold ischemia time at room temperature of 0-4 h. Similarly, the intensity of PTEN immunostaining in the second sample set did not differ significantly among benign prostate tissue subjected to cold ischemia times that ranged between 0.5 - 48 h

  • Biospecimens

    TissueProstate
  • Preservation Types

    Formalin
  • Diagnoses

    NeoplasticBenign
  • Platforms

    ProteinImmunohistochemistry
    ProteinTissue microarray
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionCold ischemia time
    0 min
    1 h
    2 h
    30 min
    4 h
    ≤ 48 h
    Immunohistochemistry SpecificTargeted peptide/protein
    PTEN