High-throughput isolation of circulating tumor DNA: a comparison of automated platforms.

van Dessel LF, Vitale SR, Helmijr JCA, Wilting SM, van der Vlugt-Daane M, Oomen-de Hoop E, Sleijfer S, Martens JWM, Jansen MPHM, Lolkema MP

Publication: Mol Oncol (2019) • Vol 13 • Page 392-402PMID: 30516338View on PubMed

Purpose

This paper investigated the effects of carrier RNA (cRNA) addition, extraction method, and blood collection tube type on the yield and fragment size of cell-free DNA (cfDNA). The effect of automated extraction method on genotyping results was also investigated.

Conclusion

The use of cRNA reduced the yield of small (136 bp) DNA without altering recovery of large DNA (? bp) and, while it increased the cfDNA yield as determined by Qubit, it did not alter yield by the PCR-based TERT assay. Overall, the CellSave tubes were found to be compatible with the automated extraction methods, but use of CellSave tubes resulted in significantly fewer 2000 bp fragments than when EDTA tubes were used, regardless of extraction kit. Further, blood from CellSave tubes produced lower cfDNA yields with the Maxwell kit and fewer copies of the 136 bp fragment when extracted using the QIAsymphony kit when compared to blood from EDTA tubes. Although overall cfDNA yields and copies of the different size fragment from the plasma of healthy patients were comparable between extraction methods, cfDNA yields from the plasma of cancer patients were higher when extraction was with QIAamp or QIAsymphony kits than the MAXwell kit. Use of the Maxwell kit resulted in fewer 136 bp fragments than the QIAamp kit and less of the 2000 bp fragment than the QIAsymphony kit. The number of wildtype and variant copies were more comparable to that in QIAamp when extraction was with QIAsymphony than the Maxwell kit. While significantly fewer wildtype copies were detected using the Maxwell than QIAsymphony kit, the variant allele frequency (VAF) was comparable between QIAsymphony and Maxwell kits.

Studies (3)

Purpose

The purpose of this study was to determine if the use of cRNA during extraction with QIAsymphony or Maxwell increased the cfDNA recovery. Blood was collected from an unspecified number of healthy donors into either two CellSave tubes or one EDTA tube and one CellSave tube. All specimens were stored at room temperature until plasma isolation. Within 24 h (EDTA specimens) or 72 h (CellSave specimens), plasma was obtained by centrifugation at 1711 x g for 10 min at room temperature followed by 12 000 x g for 10 min at room temperature, pooled, and frozen at -80˚C until extraction. cfDNA was extracted from plasma aliquots using the QIAamp circulating nucleic acid kit, QIAsymphony SP Circulating DNA Kit, or the Maxwell RSC LV ccfDNA Plasma Custom Kit. The effect of cRNA concentration was tested by adding 0.25-4 µg to plasma along with synthetic DNA from a plant. Extracted cfDNA was quantified using Qubit Quant-iT dsDNA high Sensitivity assay and the TaqMan copy number reference assay TERT. DNA fragment size was investigated by digital PCR amplification of three beta-actin fragments (136 bp, 420 bp, and 2000 bp).

Summary of Findings

Carrier RNA (cRNA) increased the cfDNA yield during extraction with QIAsymphony (P<0.001) and Maxwell RSC (P<0.001) as determined by Qubit, but this increase was not observed when quantified by the TERT qPCR assay. cRNA reduced the recovery of small (136 bp) DNA (P=0.001 for Maxwell and P<0.001 for QIAsymphony) without changing the recovery of large (420 and 2000 bp) cfDNA. cRNA increased the recovery efficiency for plant DNA (P=0.02 for Maxwell and P=0.04 for QIAsymphony) which was approximately 30% higher using the QIAsymphony than Maxwell kit, regardless of the addition of cRNA (P<0.001).

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    DNAFluorometry
    DNADigital PCR
    DNAReal-time qPCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationAnalyte isolation method
    Maxwell RSC LV ccfDNA Plasma Custom Kit
    No Carrier RNA added
    QIAsymphony SP Circulating DNA Kit
    With Carrier RNA
    Digital PCR SpecificLength of gene fragment
    136 bp
    2000 bp
    420 bp
    Fluorometry SpecificTechnology platform
    Real-time PCR