Performance comparison of blood collection tubes as liquid biopsy storage system for minimizing cfDNA contamination from genomic DNA.

Zhao Y, Li Y, Chen P, Li S, Luo J, Xia H

Publication: J Clin Lab Anal (2018) • Page e22670PMID: 30191594View on PubMed

Purpose

This paper compared genomic DNA and hemoglobin levels in plasma from blood stored for up to 14 days pre-centrifugation in K2EDTA, Streck, and Roche tubes. The effect of storing blood for three days in Roche or Streck tubes on NGS library yield and complexity was also examined.

Conclusion

As expected, genomic DNA contamination and hemolysis of plasma increased with pre-centrifugation storage of blood in K2EDTA tube but were more stable when blood was stored in Streck or Roche tubes. Importantly, the Roche tubes but not Streck tubes were effective at preventing the release of gDNA for storage durations of 14 days. Further, while hemolysis increased when blood was stored for 14 days in either Streck or Roche tubes, the magnitude of the hemolysis was less in Roche than Streck Tubes.

NGS library yield before and after capture and library complexity were comparable when blood in K2EDTA was processed within 2 h and when blood was stored in Streck or Roche tubes for 3 days before processing.

Studies (1)

Purpose

This study compared genomic DNA and hemoglobin levels in plasma from blood stored for up to 14 days pre-centrifugation in K2EDTA, Streck, and Roche tubes. The effect of storing blood for three days in Roche or Streck tubes on NGS library yield and complexity was also examined. Blood was collected from six healthy donors into K2EDTA tubes, Streck BCT, and Roche tubes and stored at room temperature for 0, 3, 7, or 14 days before centrifugation. At the allotted time, specimens were mixed gently and aliquots removed before centrifugation at 2000 x g for 10 min at 4 ˚C followed by centrifugation of the supernatant at 16000 x g for 10 min at 4˚C. The resultant plasma was stored at -80˚C. Hemolysis was determined using a spectrophotometer. Total genomic DNA in plasma was quantified by real-time PCR of L1PA2. cfDNA was extracted from plasma using the Circulating Nucleic Acid Kit, sheared using a Covaris M220, end-repaired, adapter-ligated, selected using Agencourt beads, and amplified. The library was quantified by Qubit and LabChip GXII Touch. cfDNA was sequenced using a NextSeq500 sequencer.

Summary of Findings

As expected, genomic DNA contamination of plasma increased linearly with pre-centrifugation storage of blood in K2EDTA tubes (5.35, 118.26, and 449.85-fold after 3, 7, and 14 days, respectively). In contrast, much smaller increases were noted when blood was stored for 3, 7, or 14 days in Streck (2.18, 2.87 and 11.49-fold, respectively) or Roche tubes (1.30, 2.50, and 2.73-fold, respectively). Importantly, the Roche tubes were more effective than the Streck tubes at preventing the release of gDNA for storage durations of 14 days. Further, plasma hemoglobin levels were significantly lower than in K2EDTA blood when blood was stored for up to 7 days in Streck (P=3.55 x 10-6) or Roche (P=8.80 x 10-8) tubes. Hemolysis increased when blood was stored for 14 days in either Streck or Roche tubes but the magnitude of the hemolysis was less in Roche than Streck Tubes. Importantly, the authors state that the rupture of WBCs, as indicated by genomic DNA contamination, and the rupture of RBCs, as indicated by hemolysis, were not synchronized.  NGS library yield before and after capture and library complexity were comparable when blood in K2EDTA was processed within 2 h and when blood was stored in Streck or Roche tubes for 3 days before processing.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)Streck/BCTOther Preservative
  • Diagnoses

    Normal
  • Platforms

    DNAFluorometry
    DNANext generation sequencing
    DNAReal-time qPCR
    ProteinSpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    K2EDTA tube
    Roche tube
    Streck tube
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    0 days
    14 days
    3 days
    7 days