Critically low sodium levels due to concentration gradients formed in patient samples after undergoing a freeze-thaw cycle.

Marquez CP, Petersen JR, Okorodudu AO

Publication: Clin Chim Acta (2018) • Vol 484 • Page 218-222PMID: 29807019View on PubMed

Purpose

This paper investigated the effects of sampling position within the serum layer, a single freeze-thaw cycle, post-thaw storage duration, and post-thaw mixing on the concentration of clinical chemistry analytes in serum separator tubes (SSTs).

Conclusion

The average concentration ratio (CR) for all analytes was higher in the serum obtained from just above the separator and lower in the serum from the top of the tube than in the serum obtained immediately after inversion. When specimens were freeze-thaw cycled once after storage at either temperature and thawed for 1 h or 3 h, the average concentration of most analytes was approximately 50% and 60% that in the specimen before freezing. When mixed after thaw and before analysis, the CR increased to slightly above baseline, which the authors attribute to the removal of the more dilute areas to complete the analysis at 1 and 3 h post-thaw.

Studies (2)

Purpose

This study investigated if concentration gradients exist within serum with low sodium concentrations. From each of three specimens in SSTs (collection and processing details not provided), 200 µL aliquots were removed from the top and bottom layer and analyzed. The tubes were then inverted six or more times and analyzed. Levels of sodium, chloride, potassium, blood urea nitrogen, creatinine, calcium, and glucose were determined in the aliquots as well as in the original tube after mixing using a Vitros 5600.

Summary of Findings

Concentration gradients were observed for all analytes, with the highest concentrations occurring just above the separator. The authors report that the CR for each analyte was comparable to the other analytes tested for each specimen and thus the CR averages presented are for all analytes from each specimen. The average CR in the top of the tube to that in the mixed specimen ranged from 0.62-0.82, depending on the patient, and the CR of analytes from the bottom of the tube was 1.11-1.46.

  • Biospecimens

    FluidSerum
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen heterogeneity
    Sampling at multiple levels