Skip to main content

Stability of routine biochemical analytes in whole blood and plasma/serum: focus on potassium stability from lithium heparin.

Dupuy AM, Cristol JP, Vincent B, Bargnoux AS, Mendes M, Philibert P, Klouche K, Badiou S

Publication: Clin Chem Lab Med (2018) • Vol 56 • Page 413-421PMID: 28888086View on PubMed

Purpose

This paper investigated the effects of delayed blood processing and delayed analysis on levels of 23 clinical chemistry analytes in plasma and nine analytes in serum.

Conclusion

Storage of blood for up to 24 h before centrifugation affected only the measurement of potassium (by 8 h), bicarbonate (by 12 h), lactate dehydrogenase (LDH, by 12 h), and magnesium (by 24 h) in plasma and C-peptide, osteocalcin, and parathyroid hormone (PTH) in serum. Post-centrifugation storage of plasma in the collection tube affected the measurement of LDH, bicarbonate, potassium, phosphorous, and aspartate aminotransferase (AST) in plasma and post-centrifugation storage of serum significantly affected levels of osteocalcin and PTH.

Studies (2)

Purpose

This study investigated the effects of delayed blood processing and delayed analysis on levels of 23 clinical chemistry analytes in plasma. Blood was collected from six healthy patients and four patients in the intensive care units into seven Vacuette lithium heparin tubes. Plasma was obtained by centrifugation of blood at 2000 x g for 10 min both immediately (0 h) and after 2, 4, 6, 8, or 24 h at room temperature with inversion every 20 min.  Specimens were analyzed immediately after centrifugation and again after 2, 4, 6, 8, 12, 24, and 48 h at 4˚C. Levels of bicarbonate, ferritin, LDH, magnesium, NT-proBNP, phosphorus, potassium, total bilirubin, high-sensitivity troponin T, alkaline phosphatase, alanine aminotransferase, aspartate aminotransferase, urea, calcium, chloride, creatine kinase, creatinine, C-reactive protein, gamma-glutamyl transferase, lipase, myoglobin, sodium, and uric acid were determined using a Cobas 8000 system. Instability was defined as deviation greater than the total change limit as defined in a previous publication.

Summary of Findings

Storage of blood at room temperature prior to centrifugation resulted in unacceptable changes in levels of potassium after 8 h (-6.5%), bicarbonate (-8.8%) and LDH (12.6%) after 12 h, and magnesium after 24 h (7.7%). All other analytes in plasma remained stable when blood was stored for up to 24 h before centrifugation. Post-centrifugation storage of plasma in the collection tube resulted in significant changes in LDH after 2 h (11.5%), bicarbonate after 4 h (-11.3%), and potassium and phosphorous after 24 h (9.6% and 7.7%, respectively). Levels of AST increased 14.7% when stored for 48 h post-centrifugation, but stability was unclear as earlier time-points weren’t tested. All other analytes remained unchanged after a refrigerated storage for 48 h in the centrifugation tube.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Normal
    Other diagnoses
  • Platforms

    Small moleculeClinical chemistry/auto analyzer
    PeptideClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Electrolyte/MetalClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    StorageStorage duration
    0 h
    12 h
    2 h
    24 h
    4 h
    48 h
    6 h
    8 h
    StorageTime at room temperature
    0 h
    12 h
    2 h
    24 h
    4 h
    6 h
    8 h