Pre-storage centrifugation conditions have significant impact on measured microRNA levels in biobanked EDTA plasma samples.

Binderup HG, Houlind K, Madsen JS, Brasen CL

Publication: Biochem Biophys Rep (2016) • Vol 7 • Page 195-200PMID: 28955906View on PubMed

Purpose

This paper investigated the effects of different centrifugation protocols on platelet counts and microRNA (miRNA) levels. Additionally, the differences in platelet counts due to sampling location within the plasma phase of centrifuged blood was examined.

Conclusion

While a single centrifugation step or recentrifugation at 3000 x g for 10 min after frozen storage was not sufficient to remove all platelets, platelet levels were comparable to platelet-poor plasma (PPP) in all specimens when frozen plasma was thawed and centrifuged twice for 15 min at 3000 x g or in 6 of 10 specimens when thawed and centrifuged once at 3000 x g for 30 min. Despite comparable levels of platelets, miRNA levels were significantly higher when plasma was stored frozen before platelet removal than in PPP.

Studies (1)

Purpose

This study investigated the effects of different centrifugation protocols on platelet counts and miRNA levels. Additionally, the differences in platelet counts due to sampling location within the plasma phase of centrifuged blood was examined. Blood was collected from 10 healthy patients using a 21-gauge needle into three K2EDTA tubes. One tube from each patient was centrifuged at 3000 x g for 15 min and the resultant plasma was  then centrifuged at 3000 x g for 15 min to produce platelet-poor plasma (PPP) which was stored at -80°C. Another tube was centrifuged for 10 min at 2000 x g and the resultant plasma was frozen for a week at -80°C. Following thaw, the plasma was split and half was centrifuged at 3000 x g for 15 min to remove residual platelets, assayed, and then centrifuged again at 3000 x g for 15 min before reanalysis while the other half was analyzed, centrifuged at 3000 x g for 30 min, and then analyzed again. The third tube was centrifuged for 10 min at 2000 x g, the resultant plasma was transferred in 1 mL aliquots from the top down, and then each aliquot was analyzed separately. miRNA was isolated using the Nucleospin miRNA kit and reverse transcribed using the TaqMan MicroRNA Reverse Transcription Kit and Custom TaqMan MIR RT Pool and then quantified using custom TaqMan low density arrays. miRNA and cDNA were frozen at -20°C until use. Platelets were counted using a hematology autoanalyzer. 

Summary of Findings

As expected, the highest platelet counts were observed in specimens subjected to a single centrifugation step (2000 x g for 10 min or 3000 x g for 15 min). Although residual platelets were observed in specimens subjected to centrifugation at 2000 x 10 min followed by centrifugation at 3000 x g for 15 min, the amount decreased to 0-1x10^9 (equivalent to PPP) for all specimens when centrifuged again for 15 min at 3000 x g or in 6 of 10 specimens when the second centrifugation was extended to 30 min.

In specimens subjected to a single centrifugation for 10 min at 2000 x g, platelets were generally equally distributed throughout the tube but the lowest aliquot contained a higher number of platelets in two of the ten patients.

Compared to PPP, plasma obtained by centrifugation at 2000 x g for 10 min with or without an additional centrifugation step after frozen storage had higher levels of miR-142-3p (4.7-37.3 fold, P<0.05), miR-145 (4.9-68.4 fold, P<0.05), miR-26a (2.4-92.3-fold, P<0.05), miR-28 (9.3-102.5-fold, P<0.05), miR-301 (11.8-84-fold, P<0.05), miR-30a-5p (2.5-9.8-fold, P<0.05), miR30-d (2.8-13.9-fold, P<0.05), miR-328 (23.0-94.1 fold, P<0.05), miR-331 (8.4-113.1-fold, P<0.05), miR-335 (17.1-95.1, P<0.05), miR-340 (8.3-53..7-fold, P<0.05), miR-92a (2.5-13.8 fold, P<0.05), and miR-93 (3.5-13.1 fold, P<0.05). With the exception of miR-26a, significant differences were observed in miR for all specimens centrifuged twice at 3000 x g for 15 min after frozen storage compared to PPP (P<0.001). When normalized to the exogenous control cel-miR-39, levels of miR-16 were found to be significantly higher in specimens stored at -80°C after an initial centrifugation at 2000 x g for 10 min than in PPP (1.6 to 2.6-fold, P<0.05, all). Consequently, levels of miR normalized to miR-16 showed smaller fold changes between protocols but most differences remained significant.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    Frozen
  • Diagnoses

    Normal
  • Platforms

    RNAReal-time qRT-PCR
    Cell count/volumeHematology/ auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsBiospecimen heterogeneity
    Sampling at multiple levels
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    Different number of centrifugation steps compared
    Multiple durations compared
    Multiple speeds compared
    StorageStorage conditions
    As PPP
    As plasma
    Real-time qRT-PCR SpecificTargeted nucleic acid
    miR-142-3p
    miR-145
    miR-16
    miR-26a
    miR-28
    miR-301
    miR-30a-5p
    miR-30d
    miR-328
    miR-331
    miR-335
    miR-340
    miR-92a
    miR-93