This paper investigated the effects of pneumatic tube system (PTS) transport of blood on hematology and clinical chemistry analytes.
Effects of one directional pneumatic tube system on routine hematology and chemistry parameters; A validation study at a tertiary care hospital.
Lee AJ, Suk Suh H, Jeon CH, Kim SG
Purpose
Conclusion
Significant effects of PTS were observed on aspartate aminotransferase (AST), lactate dehydrogenase (LDH), hemolysis index, and mean platelet volume (MPV) when compared to hand-delivered specimens, but all values remained within the 95% confidence interval with the exception of MPV for two specimens and potassium levels for three specimens. All other analytes were unaffected by PTS transport.
Studies (1)
Purpose
This study investigated the effects of PTS transport on 18 hematology and 20 clinical chemistry analytes. Blood specimens from 57 preoperative patients (diagnosis not specified) were collected into two vacutainer K2EDTA tubes and two vacutainer SST tubes during a single venipuncture. One tube of each type was transported 347 feet (including four stories of vertical rise) by PTS at a speed of 7-10 m/sec with acceleration and deceleration at the stations. The matched set of tubes was hand-delivered. Specimens were analyzed in parallel. The complete blood count with WBC differential and platelet parameters was determined in EDTA anticoagulated whole blood using a hematology autoanalyzer. Chemistry parameters were determined in serum specimens using a Cobas 8000.
Summary of Findings
Compared to hand-delivered specimens, specimens delivered by PTS had higher AST, LDH, and hemolysis index (P=0.000, all). While there was a clear bias, the effect was small and values for only 2-3 of the 57 specimens fell outside of the 95% confidence interval. Although there was a significant difference in MPV between PTS and hand-delivered specimens (P=0.001), the mean values (8) and ranges were identical (7.5-8.5) and with the exception of two outliers, all values remained within the 95% confidence interval. Potassium values were comparable in PTS and hand-delivered specimens, but the values fell outside the 95% confidence interval after PTS for three specimens (two high, one low). There was no effect of PTS on red blood cell (RBC), white blood cell (WBC), and platelet counts, hemoglobin, hematocrit, mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), MCH concentration (MCHC), red cell distribution width (RDW), erythrocyte sedimentation rate (ESR), percentage neutrophils, lymphocytes, monocytes, eosinophils and basophils, the absolute neutrophil count, platelet volume distribution width (PDW), alkaline phosphatase (ALP), γ -glutamyl transferase (GGT), total bilirubin, cholesterol, triglyceride, glucose, protein, albumin, blood urea nitrogen (BUN), creatinine, uric acid, sodium, chloride, lipemia index, or icterus index.
Biospecimens
FluidBloodFluidSerumPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
CarbohydrateClinical chemistry/auto analyzerCell count/volumeHematology/ auto analyzerElectrolyte/MetalClinical chemistry/auto analyzerLipidClinical chemistry/auto analyzerMorphologyHematology/ auto analyzerProteinClinical chemistry/auto analyzerProteinHematology/ auto analyzerSmall moleculeClinical chemistry/auto analyzerSteroidClinical chemistry/auto analyzerPre-analytical Factors