Evaluation of a reduced centrifugation time and higher centrifugal force on various general chemistry and immunochemistry analytes in plasma and serum.

Møller MF, Søndergaard TR, Kristensen HT, Münster AB

Publication: Ann Clin Biochem (2017) • Vol 54 • Page 593-600PMID: 27687083View on PubMed

Purpose

This paper investigated the effects of centrifuging blood for 5 min at 3000 x g rather than the standard of 10 min at 2200 x g on levels of nine clinical chemistry analytes in lithium heparin plasma and serum.

Conclusion

With the exception of lactate dehydrogenase (LDH) in lithium heparin plasma, the mean bias and coefficient of variance (CV) of all analytes remained within the allowable bias after centrifugation at 3000 x g for 5 min rather than the standard protocol. For LDH, the faster and shorter centrifugation resulted in an increase in the coefficient of variance.  In specimens from two patients, the faster and shorter centrifugation resulted in a higher than acceptable bias in protein and free triiodothyronine (FT3) levels. Additionally, two plasma specimens had higher than acceptable hemolysis index after the faster and shorter centrifugation and one serum specimen had higher than acceptable hemolysis index using the standard centrifugation.

Studies (2)

Purpose

This study investigated the effects of centrifuging blood for 5 min at 3000 x g rather than the standard protocol of 10 min at 2200 x g on levels of nine clinical chemistry analytes in lithium heparin plasma. Blood specimens from 40 patients and healthy volunteers were collected in quadruplicate into lithium heparin vacutainer tubes. Within 1 h of venipuncture, two specimens from each patient were centrifuged at 21˚C for 5 min at 3000 x g while the remaining two were centrifuged using the standard protocol of 10 min at 2200 x g. The same acceleration and braking speeds (step 9, 25 s) were used in conjunction with both centrifugation protocols. Levels of calcium (Ca), immunoglobulin M (IgM), phosphorus (P), potassium (K), hemolysis index (Hem), icterus index (ICT), lipaemia index (LIP), and LDH were determined using an Abbott Architect c16000 and levels of cobalamin (vitamin B12), folic acid (FOL), and FT3 were determined using an Abbott Architect i2000SR.

Summary of Findings

Lithium heparin plasma obtained by centrifugation at 3000 x g for 5 min rather than the standard protocol had higher average levels and an increase in the coefficient of variance of LDH. The average bias in LDH from the alternate protocol was 6.3%, which exceeded the goal of 4.3%. The mean bias and CV of all other analytes remained within the allowable bias after centrifugation at 3000 x g for 5 min rather than the standard protocol. However, the difference in FT3 levels were higher than acceptable in specimens from two patients and the hemolysis index was greater than acceptable for two other patients after centrifugation at 3000 x g for 5 min.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
    Normal
  • Platforms

    Electrolyte/MetalClinical chemistry/auto analyzer
    GlycoproteinClinical chemistry/auto analyzer
    PeptideChemiluminescence immunoassay
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen Aliquots and ComponentsCentrifugation
    Multiple durations compared
    Multiple speeds compared