Various glycolysis inhibitor-containing tubes for glucose measurement cannot be used interchangeably due to clinically unacceptable biases between them.

Saracevic A, Dukic L, Juricic G, Milevoj Kopcinovic L, Mirosevic G, Simundic AM

Publication: Clin Chem Lab Med (2018) • Vol 56(2) • Page 236-241.PMID: 28746044View on PubMed

Purpose

This paper investigated the effects of anticoagulant and collection tube type on the measurement of glucose in blood from healthy volunteers and patients with diabetes.

Conclusion

Although levels of glucose were comparable between specimens collected in lithium heparin tubes and NaF/KOx tubes, they were significantly higher in specimens collected in citrate buffer-containing tubes (Glucomedics, GlucoExact, and FC mix) and differed significantly among the citrate tubes.  Importantly, the bias compared to lithium heparin blood exceeded the specified maximum bias when collected into any of the three citrate buffer-containing tubes, regardless of patient diabetes status. While adjusting the multiplication factor decreased the bias for specimens in GlucoExact tubes to a clinically acceptable level, the bias remained clinically unacceptable after adjusting the multiplication factor for specimens collected in Glucomedics tubes.

Studies (1)

Purpose

This study investigated the effect of anticoagulant and collection tube type on the measurement of glucose in blood from twenty healthy volunteers and twenty patients with diabetes. Blood was drawn first into the lithium heparin tube and then in random order into an FX NaF/KOx tube, Glucomedics tube containing EDTA/ sodium flouride/citric acid/sodium citrate), FC mix tube containing sodium EDTA/sodium flouride/ citric acid/ sodium citrate, and S-Monovette GlucoEXACT tube containing sodium flouride /citrate/EDTA. After collection, all specimens were mixed following the appropriate tube instructions and the LiH tube was placed in an ice slurry. Within 15 min of collection, blood was centrifuged at 1800 x g for 10 min at 4˚C. Glucose concentrations were measured within 30 min of collection using the hexokinase method on an Architect c8000 analyzer. Glucose values for specimens collected in the Glucomedics and GlucoEXACT tube were multiplied by 1.16 as suggested by the manufacturer. 

Summary of Findings

Although levels of glucose were comparable between specimens collected in lithium heparin tubes and NaF/KOx tubes, they were significantly different from those in the citrate buffer-containing tubes (Glucomedics, GlucoExact, and FC mix) and differed significantly among the citrate tubes, regardless of patient diabetes status. Significantly higher glucose levels were found in specimens collected in Glucomedics, GlucoExact, and FC mix compared to specimens collected in the lithium heparin tube (7.3%, 3.2%, and 2.0%, respectively), those from patients with diabetes (7.1%, 3.2%, and 1.4%, respectively) and those from healthy individuals (7.5%, 3.1%, and 2.6%, respectively). Importantly, the bias compared to lithium heparin blood exceeded the specified maximum bias when collected into any of the three citrate buffer-containing tubes and the bias between specimens collected in the Glucomedics and GlucoExact or FC mix tubes exceeded the specified maximum bias. Adjusting the multiplication factor from 1.16 to 1.12 decreased the bias for specimens in GlucoExact tubes to -0.1%, which is clinically acceptable, but the bias only decreased to 3.2% (clinically unacceptable) for specimens collected in Glucomedics tubes. 

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Diabetes Type 1
    Diabetes Type 2
    Normal
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionAnticoagulant
    Citrate/sodium fluoride EDTA
    Lithium heparin
    Multiple manufacturers evaluated
    Sodium fluoride/potassium oxalate
    Biospecimen AcquisitionType of collection container/solution
    FC mix tube
    GlucoExact tube
    Glucomedics tube
    Lithium heparin tube
    NaF/KOx tube
    Clinical chemistry/auto analyzer SpecificData handling
    Values multiplied by 1.12
    Values multiplied by 1.16