Performance of Streck cfDNA Blood Collection Tubes for Liquid Biopsy Testing.

Medina Diaz I, Nocon A, Mehnert DH, Fredebohm J, Diehl F, Holtrup F

Publication: PLoS One (2016) • Vol 11 • Page e0166354PMID: 27832189View on PubMed

Purpose

This paper investigated the effects of storing blood in Streck blood collection tubes (BCT) at room temperature on cell-free DNA (cfDNA) yield, genomic DNA contamination, and mutation detection in plasma from healthy volunteers and patients with colorectal cancer. The effects of storage of blood at extreme temperatures (4-40˚C) on plasma volume and genomic DNA contamination was also investigated

Conclusion

The real-time PCR assay revealed approximately 20% of the DNA in plasma was ≥ 402 bp (likely genomic) when blood was stored in K2EDTA or BCT tubes for 2 h at room temperature. cfDNA yields, ratio of long to short LINE-1 PCR products, mutation rates by BEAMing, and number of each type of base change by SafeSeq were comparable when blood was stored in a BCT for up to 5 days at room temperature to those when blood was stored for 2 h in a K2EDTA tube. There was also no effect when blood was stored with agitation in a BCT. The authors report no observable effect of BCT storage on the electrophoretic migration of the DNA. However, storage of blood in BCT tubes at extreme temperatures (4, 6, 10 or 40˚C) resulted in a decrease in plasma volume and in the case of storage at 4 or 40˚C for 5 days or 6˚C for 5 days an increase in the ratio of long to short LINE-1 PCR products indicating an increase in genomic DNA contamination. Slight changes in the average mutant fraction of EGFR and KRAS, but not PIK3CA occurred when blood spiked with DNA carrying mutations was stored for 5 days in BCT tubes rather than 2h in K2EDTA or BCT tubes, but the effect was eliminated when each specimen was normalized to its respective 2 h K2EDTA control and significance of the effect was not established.  Mutation rates were strongly to very strongly correlated between plasma from blood stored for 2 h in K2EDTA tubes and that stored for 2 h or 3 days in BCT tubes in patients with colorectal cancer. 

Studies (2)

Purpose

This study investigated the effects of storing blood in Streck blood collection tubes (BCT) at room temperature on cell-free DNA (cfDNA) yield, genomic DNA contamination, and mutation detection in plasma from healthy volunteers and patients with colorectal cancer. Venous blood was collected into K2EDTA (10 mL) and BCT tubes (10 mL) and inverted 10 times before storage. Specimens from 60 healthy donors were stored in K2EDTA tubes for 2 h at room temperature and in BCT tubes for 2 h, 3 days (with or without shaking), and 5 days at room temperature before plasma isolation and DNA quantification and mutation analysis with BEAMing and Safe-SeqS. Specimens from 21 patients with CRC and specimens from eight healthy donors spiked with DNA carrying specific mutations were stored in K2EDTA tubes for 2 h at room temperature and in BCT tubes for 2 h and 3 days at room temperature before plasma isolation and DNA quantification and mutation analysis with BEAMing. Plasma was obtained by centrifugation for 10 min at 1600 x g followed by centrifugation of the supernatant at 6000 x g for 10 min, aliquoted, and frozen at -80˚C. Plasma was thawed at room temperature and DNA was extracted using the QIAamp Circulating Nucleic Acid Kit with an extension of the Proteinase K digestion step to 1 h. DNA was quantified by real-time PCR amplification of 96 bp (cfDNA and gDNA) and 402 bp (gDNA only) of LINE-1. Mutation analysis was conducted using the flow cytometry-based BEAMing method and the next-generation sequencing based SafeSeqS.

Summary of Findings

The real-time PCR assay revealed approximately 20% of the DNA in plasma was ≥ 402 bp (a size likely to represent genomic DNA) when blood was stored in K2EDTA or BCT tubes for 2 h at room temperature. Comparable plasma yields of cfDNA, ratios of long to short LINE-1 PCR products, mutation rates by BEAMing, and number of each type of base changes by SafeSeq were obtained from blood stored in BCT or EDTA tubes for 2 h at room temperature and from blood stored in a BCT at room temperature for 3 days with or without shaking or stored for 5 days. The authors state that the electrophoretic migration was also comparable for plasma obtained from blood stored in BCT and K2EDTA tubes. A slight increase in the average mutant fraction of EGFR (0.6% versus 0.4%) and a decrease in the mutant fraction of KRAS (0.6% versus 0.7%) occurred when blood spiked with DNA carrying mutations was stored for 5 days in BCT tubes rather than 2h in K2EDTA or BCT tubes, but no change in the average mutant fraction of PIK3CA was observed and it was unclear if the effect was statistically significant or was within the expected variability. Further, analysis showed than when individual values were normalized to their respective K2EDTA control there was no difference in fold change. Plasma DNA yields were comparable and mutation rates were strongly to very strongly correlated (R2=0.876 to 0.973) between plasma from blood stored for 2h in K2EDTA tubes and that stored for 2 h or 3 days in BCT tubes in patients with colorectal cancer. 

  • Biospecimens

    FluidPlasma
    FluidBlood
  • Preservation Types

    Streck/BCTNone (Fresh)
  • Diagnoses

    Normal
    NeoplasticCarcinoma
  • Platforms

    DNAReal-time qPCR
    DNANext generation sequencing
    DNAFlow cytometry
    DNAElectrophoresis
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    K2EDTA tube
    Streck BCT
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    Biospecimen PreservationType of fixation/preservation
    Blood collection tube additive
    EDTA
    StorageStorage conditions
    With agitation
    Without agitation
    StorageTime at room temperature
    2 h
    3 days
    5 days