Study of Preanalytic and Analytic Variables for Clinical Next-Generation Sequencing of Circulating Cell-Free Nucleic Acid.

Mehrotra M, Singh RR, Chen W, Huang RSP, Almohammedsalim AA, Barkoh BA, Simien CM, Hernandez M, Behrens C, Patel KP, Routbort MJ, Broaddus RR, Medeiros LJ, Wistuba II, Kopetz S, Luthra R

Publication: J Mol Diagn (2017) • Vol 19(4) • Page 514-24PMID: 28506684View on PubMed

Purpose

This paper investigated the effects of blood collection tube type and delayed centrifugation on the yield and size of cell-free DNA (cfDNA). The mutational status obtained with cfDNA and formalin-fixed paraffin-embedded FFPE tumor specimens were also compared for a given patient. Optimization of next generation sequencing (NGS) using different platforms was also discussed.

Conclusion

Yields of cfDNA were comparable between specimens collected in Streck blood collection tubes (BCT) and K3-EDTA tubes when plasma was isolated within 16 h of blood collection, but significantly more cfDNA was obtained from specimens collected in K3-EDTA than BCT tubes when centrifugation was delayed by >24 h.. cFDNA from specimens collected in K3-EDTA tubes also had a higher ratio of larger fragments than specimens collected in BCT tubes. Importantly, there was 100% concordance in mutant allele frequencies by Ion PGM Sequencing when specimens collected in Streck and K3-EDTA tubes were compared.

Of the 54 mutations that were expected based on sequencing data obtained with FFPE tissue, 45 (87%) were also identified when plasma specimens were sequenced. Upon manual inspection, all mutations found in the tissue specimens were confirmed in the plasma specimens, but they occurred at an allele frequency below the 1% allele frequency threshold. In addition, two mutations were identified in plasma but not tissue specimens using the Ion Proton and the MiSeq platforms, but were not confirmed by manual inspection of reads obtained with tissue specimens.

When 5 pmol DNA was used for library construction (rather than 20 pmol DNA), the number of pyrosequencing reads was higher and the percentage of polyclonal reads was lower, but the variant allele frequency was unaffected by input amount. Pyrosequencing using the Ion Proton platform, as opposed to the Ion PGM platform, resulted in 15.7-fold more reads and 2.3-fold greater mean depth, but allelic calls were 100% concordant between the two platforms. Concordance between MiSeq and Ion Proton results was observed for 50 of the 52 variants detected in plasma. The mutations identified by MiSeq differed slightly when an alternate bioinformatics pipeline was used.

Studies (3)

Purpose

This study investigated the effects of blood collection tube type and delayed centrifugation on the yield and size of cfDNA, as well as the detection of mutations using next-generation sequencing. Two methods of DNA quantification were also compared. The study included peripheral blood from 31 patients with known mutational status including eight patients with endometrial adenocarcinomas, seven with colon adenocarcinomas, three with breast adenocarcinomas, three with prostate adenocarcinomas, one with an ovarian carcinoma, one with a lung adenocarcinoma, one with a liver adenocarcinoma, one with an esophagus signet ring adenocarcinoma, one with an appendiceal mucinous adenocarcinoma, one with a squamous cell carcinoma of the tongue, one with a parathyroid carcinoma, and three with astrocytoma or ganglioneuroblastoma. Blood was drawn into both a regular K3-EDTA tube and a cell-free DNA BCT tube. Both tubes were inverted 10 times, stored for 0, 2, 4, 16 and >24 h and centrifuged at 2000 x g for 10 min at room temperature. Plasma was transferred to a new tube and centrifuged again at 2000 x g for 10 min at room temperature before extraction of cfDNA using the QIAamp Circulating Nucleic Acid Kit. cfDNA was stored at -20˚C until analysis. DNA was quantified using the Qubit dsDNA HS Assay and by amplifying 115 and 247 bp Alu repeats. Four cfDNA specimens were also sequenced using the Ion PGM.

Summary of Findings

High molecular weight DNA was not present in plasma specimens as determined by bioanalyzer when centrifuged within 16 h regardless of the collection tube type. cfDNA yield was strongly correlated between blood collection tube types (r=0.820, P<0.0001) and values were comparable when plasma was obtained within 16 h of blood collection. However, significantly more cfDNA was obtained from specimens collected in K3-EDTA than BCT tubes when centrifugation was delayed by more than 24 h, (37.03 ng/mL versus 22 ng/mL , P<0.0001). cfDNA isolated from specimens collected in K3-EDTA tubes also had a higher ratio of real-time PCR amplicons (Alu 247 to Alu 115) than specimens collected in BCT tubes (0.39-0.55 versus 0.35-0.43, P<0.01). Importantly, when plasma specimens collected in Streck and K3-EDTA tubes were sequenced by Ion PGM mutant allele frequencies were 100% concordance and the variant allele frequencies (VAFS) were very strongly correlated (r=0.95, P<0.0001). Calculated DNA yields were slightly higher when determined by Qubit then by Alu real-time PCR.

  • Biospecimens

    FluidBlood
    FluidPlasma
  • Preservation Types

    None (Fresh)Streck/BCT
  • Diagnoses

    NeoplasticCarcinoma
    NeoplasticOther
  • Platforms

    DNANext generation sequencing
    DNAFluorometry
    DNAAutomated electrophoresis/Bioanalyzer
    DNAReal-time qPCR
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen AcquisitionType of collection container/solution
    K3-EDTA Vacutainer
    Streck BCT
    Biospecimen Aliquots and ComponentsCentrifugation
    Centrifugation delays investigated
    Biospecimen PreservationType of fixation/preservation
    Blood collection tube additive
    None (fresh)
    StorageTime at room temperature
    0 h
    16 h
    2 h
    4 h
    >24 h
    Real-time qPCR SpecificLength of gene fragment
    115 bp
    247 bp
    Real-time qPCR SpecificTargeted nucleic acid
    Alu
    Fluorometry SpecificTechnology platform
    Real-time PCR