RNA Extracted from Formalin-Fixed, Paraffin-Embedded Renal Biopsy Biospecimens: An Evaluation of Alternative Extraction Kits and the Effects of Storage Time.

Lu Y, Zhang M, Jiang Q, Yin R, Zhu P, Wang H, Liu Z, Zheng C

Publication: Biopreserv Biobank (2017) • Vol 15(4) • Page 396-398PMID: 28426236View on PubMed

Purpose

This paper compared yield, purity, and integrity of RNA extracted using two different methods from formalin-fixed paraffin-embedded (FFPE) renal biopsy specimens and between specimens stored for 5 and 10 years.

Conclusion

Extraction with the RecoverAll kit resulted in more amplifiable RNA and more intact RNA than with the RNeasy FFPE kit as determined by real-time PCR, but the RNA integrity numbers (RIN) and purity were comparable between the two extraction methods. RNA yield was significantly lower from specimens stored for 10 years rather than 5 years, but had comparable RIN numbers, purity, and number of amplifiable copies of 18S rRNA, CTNNB1, and SYNPO.

Studies (2)

Purpose

This study compared the yield, purity, and integrity of RNA obtained from FFPE renal biopsy specimens using two different extraction kits. Renal biopsies (<1.2mm by 10mm) from twelve patients with IgA nephropathy were fixed for 4-6 h, paraffin-embedded, and stored at room temperature in the dark for 5 years. RNA was extracted from a single 5 µm section using both the RecoverAll Total Nucleic Acid Isolation kit for FFPE and RNeasy FFPE kit. RNA yield was determined spectrophotometrically, RNA integrity was assessed by bioanalyzer, and 86 bp and 136 bp amplicons of 18S rRNA were analyzed by real-time PCR.

Summary of Findings

Slightly but not significantly higher RNA yields were obtained when RNA was extracted using the RecoverAll kit than the RNeasy kit (261 vs. 159 ng/5 mm section). The mean RIN values and purity were comparable in specimens extracted using RecoverAll and RNeasy (2.3 versus 2.4 and A260/A280 of 1.54 versus 1.50) and most of the extracted RNA fragments were <200nt, regardless of extraction method used.  However, the CT values for both 18S rRNA amplicons were higher when RNA was extracted using RNeasy rather than RecoverAll (21.66 versus 22.71, P<0.03 for the 86 bp fragment and 24.42 versus 22.02, P<0.001for the 136 bp fragment). Further, the difference in the CT values for the 136 bp fragment and the 86 bp fragment was lower for specimens extracted using the RecoverAll kit than the RNeasy kit (0.36 versus 1.71, P=0.03), confirming that the RNA obtained using the RecoverAll kit was more intact.

  • Biospecimens

    TissueKidney
  • Preservation Types

    Formalin
  • Diagnoses

    Other diagnoses
  • Platforms

    RNAAutomated electrophoresis/Bioanalyzer
    RNAReal-time qRT-PCR
    RNASpectrophotometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Analyte Extraction and PurificationAnalyte isolation method
    RNeasy FFPE kit
    RecoverAll Total Nucleic Acid Isolation Kit
    Real-time qRT-PCR SpecificLength of gene fragment
    136 bp
    86 bp
    Real-time qRT-PCR SpecificTargeted nucleic acid
    18S rRNA
    Real-time qRT-PCR SpecificTechnology platform
    Bioanalyzer