This paper compared albumin levels and albumin creatinine ratios (ACR) among urine specimens stored on pantyliners or PeeSpots for 3 h at 37˚C followed by 72 h at room temperature and those stored in tubes refrigerated for 75 h.
Comparison of urine collection methods for albuminuria assessment in young children.
van den Belt SM, Gracchi V, de Zeeuw D, Heerspink HJ
Purpose
Conclusion
Urine stored on PeeSpots had comparable levels of albumin and ACR to matched refrigerated specimens, but significantly lower concentrations were observed for urine stored on pantyliners. Compared to the PeeSpots, use of a pantyliner resulted in higher bias and precision but lower accuracy for albumin and ACR. Although the sensitivity for detection of urinary albumin concentration of more than 20 mg/L was higher for specimens collected on PeeSpots than pantyliners, the specificity was higher for specimens collected on pantyliners.
Studies (1)
Purpose
This study compared albumin levels and ACR among urine specimens stored on pantyliners or on PeeSpots for 3 h at 37˚C followed by 72 h at room temperature and specimens stored for 75 h in a refrigerator. Urine from 81 patients was pipetted onto a pantyliner and a PeeSpot and incubated at 37˚C for 3 h and then stored at room temperature for 72 h (to simulate shipping). The remainder of the urine specimen was stored refrigerated in a tube for 75 h. After 72 h, the pantyliner was opened and the wool was placed in a tube and centrifuged at 10,000 rpm for 10 min. The PeeSpot filter was placed directly in a tube and centrifuged at 350 x g for 5 min. Urine was then pipetted into tubes and albumin and creatinine levels were determined using a Roche Modular P.
Summary of Findings
Compared to the refrigerated specimen, urine stored on pantyliners had lower concentrations of urinary albumin (32.0 mg/L versus 66.0 mg/L, P<0.001) and lower albumin creatinine ratio (ACR) (8.8 mg/mm versus 23.3 mg/mm, P<0.001). In contrast, urine stored on PeeSpots had comparable levels of albumin (61.0 mg/L versus 66.0 mg/L, P=0.084) and ACR (23.9 mg/mm versus 23.3 mg/mm, P=0.53) to refrigerated specimens. Importantly, use of a pantyliner resulted in higher bias (-34.2% versus 3.3%) and precision (71.2% versus 4.3%) and lower accuracy (48.1% versus 96.3%) for albumin compared to the PeeSpot. For the ACR, use of a pantyliner instead of a PeeSpot resulted in higher bias (-33.3% versus 3.4%) and precision (74.8% versus 6.3%) but lower accuracy (46.9% versus 92.6%). Specificity for detection of urinary albumin concentrations of more than 20 mg/L was higher in specimens collected on a pantyliner than specimens collected on PeeSpots (100% versus 93.3%, respectively) but sensitivity was lower (64.2% versus 100%, respectively).
Biospecimens
FluidUrinePreservation Types
Other PreservativeDiagnoses
Not specifiedPlatforms
ProteinClinical chemistry/auto analyzerPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Storage Storage conditions On PeeSpotOn pantylinerRefrigerated in tubeStorage Storage temperature 3 h at 37˚C and 72 at room temperature75 h refrigerated