Skip to main content

Fragmentation of DNA affects the accuracy of the DNA quantitation by the commonly used methods.

Sedlackova T, Repiska G, Celec P, Szemes T, Minarik G

Publication: Biol Proced Online (2013) • Vol 15 • Page 5PMID: 23406353View on PubMed

Purpose

This paper investigated the effect of DNA fragmentation on the quantification of diluted DNA by spectrophotometric, fluorometric, and PCR-based methods.

Conclusion

Effects of DNA fragmentation on DNA quantification varied between the three methods. While DNA dilution eliminated the few effects of fragmentation observed on spectrophotometric quantification it did not change the effects on PicoGreen and real-time PCR-based quantification. Flourometric-based quantification declined with increasing fragmentation but real-time PCR was only affected when the DNA was fragmented to 150 bp.

Studies (1)

Purpose

This study investigated the effect of DNA fragmentation on the quantification of DNA by spectrophotometric, fluorometric, and PCR based methods. DNA was extracted from K2EDTA blood from 10 volunteers using the QIAamp DSP DNA Blood Mini Kit. DNA was fragmented by ultrasound to achieve specimens with target peaks at 1500 bp, 500 bp, and 150 bp and fragmentation was verified by electrophoresis. DNA was quantified by duplicate measurements using a NanoDrop spectrophotometer, PicoGreen Quant-iT and dsDNA fluoremetry, and real-time PCR amplification of a 115 bp fragment of the Alu element.

Summary of Findings

DNA yield as determined by spectrophotometer was not affected when DNA was fragmented to 150 bp but declined significantly when fragmented to 1500 bp or 500 bp (P<0.0001, both) although the decline was no longer significant when the specimen was diluted 10-fold prior to quantification. The yield of DNA declined with increasing DNA fragmentation as determined by PicoGreen, regardless of DNA dilution (10, 100, or 1000-fold). Importantly, all fragmented specimens had significantly lower amounts of DNA as determined by PicoGreen than unfragmented specimens (P<0.0001, all) except the 10-fold dilution of the DNA fragmented to 1500 bp. In contrast, fragmentation of the DNA to 150 bp resulted in decreased DNA quantification by real-time PCR, regardless of dilution (10, 100 and 1000-fold), but no effects were observed when DNA was fragmented to 1500 bp or 500 bp.

  • Biospecimens

    FluidBlood
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    DNAElectrophoresis
    DNAReal-time qPCR
    DNASpectrophotometry
    DNAFluorometry
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Real-time qPCR SpecificTargeted nucleic acid
    Alu element
    Real-time qPCR SpecificTemplate modification
    Fragmented to 150 bp
    Fragmented to 1500 bp
    Fragmented to 500 bp
    Unfragmented
    Real-time qPCR SpecificTemplate/input amount
    Diluted 1:10
    Diluted 1:100
    Diluted 1:1000
    Spectrophotometry SpecificTechnology platform
    PicoGreen
    Real-time PCR
    Spectrophotometry SpecificTemplate modification
    Fragmented to 150 bp
    Fragmented to 1500 bp
    Fragmented to 500 bp
    Unfragmented
    Spectrophotometry SpecificTemplate/input amount
    Diluted 1:10
    Undiluted
    Fluorometry SpecificTemplate modification
    Fragmented to 150 bp
    Fragmented to 1500 bp
    Fragmented to 500 bp
    Unfragmented
    Fluorometry SpecificTemplate/input amount
    Diluted 1:10
    Diluted 1:100
    Diluted 1:1000