This paper investigated the effect of DNA fragmentation on the quantification of diluted DNA by spectrophotometric, fluorometric, and PCR-based methods.
Fragmentation of DNA affects the accuracy of the DNA quantitation by the commonly used methods.
Sedlackova T, Repiska G, Celec P, Szemes T, Minarik G
Purpose
Conclusion
Effects of DNA fragmentation on DNA quantification varied between the three methods. While DNA dilution eliminated the few effects of fragmentation observed on spectrophotometric quantification it did not change the effects on PicoGreen and real-time PCR-based quantification. Flourometric-based quantification declined with increasing fragmentation but real-time PCR was only affected when the DNA was fragmented to 150 bp.
Studies (1)
Purpose
This study investigated the effect of DNA fragmentation on the quantification of DNA by spectrophotometric, fluorometric, and PCR based methods. DNA was extracted from K2EDTA blood from 10 volunteers using the QIAamp DSP DNA Blood Mini Kit. DNA was fragmented by ultrasound to achieve specimens with target peaks at 1500 bp, 500 bp, and 150 bp and fragmentation was verified by electrophoresis. DNA was quantified by duplicate measurements using a NanoDrop spectrophotometer, PicoGreen Quant-iT and dsDNA fluoremetry, and real-time PCR amplification of a 115 bp fragment of the Alu element.
Summary of Findings
DNA yield as determined by spectrophotometer was not affected when DNA was fragmented to 150 bp but declined significantly when fragmented to 1500 bp or 500 bp (P<0.0001, both) although the decline was no longer significant when the specimen was diluted 10-fold prior to quantification. The yield of DNA declined with increasing DNA fragmentation as determined by PicoGreen, regardless of DNA dilution (10, 100, or 1000-fold). Importantly, all fragmented specimens had significantly lower amounts of DNA as determined by PicoGreen than unfragmented specimens (P<0.0001, all) except the 10-fold dilution of the DNA fragmented to 1500 bp. In contrast, fragmentation of the DNA to 150 bp resulted in decreased DNA quantification by real-time PCR, regardless of dilution (10, 100 and 1000-fold), but no effects were observed when DNA was fragmented to 1500 bp or 500 bp.
Biospecimens
FluidBloodPreservation Types
None (Fresh)Diagnoses
Not specifiedPlatforms
DNAElectrophoresisDNAReal-time qPCRDNASpectrophotometryDNAFluorometryPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Real-time qPCR Specific Targeted nucleic acid Alu elementReal-time qPCR Specific Template modification Fragmented to 150 bpFragmented to 1500 bpFragmented to 500 bpUnfragmentedReal-time qPCR Specific Template/input amount Diluted 1:10Diluted 1:100Diluted 1:1000Spectrophotometry Specific Technology platform PicoGreenReal-time PCRSpectrophotometry Specific Template modification Fragmented to 150 bpFragmented to 1500 bpFragmented to 500 bpUnfragmentedSpectrophotometry Specific Template/input amount Diluted 1:10UndilutedFluorometry Specific Template modification Fragmented to 150 bpFragmented to 1500 bpFragmented to 500 bpUnfragmentedFluorometry Specific Template/input amount Diluted 1:10Diluted 1:100Diluted 1:1000