Use of the Becton-Dickinson urine culture tube with the Abbott MS-2 urine screening system.

Southern PM Jr, Luttrell B

Publication: Diagn Microbiol Infect Dis (1984) • Vol 2 • Page 193-8PMID: 6378504View on PubMed

Purpose

This paper investigated the effects of transporting and storing specimens in Becton-Dickinson (BD) tubes rather than transporting in conventional tubes on results of urine culture and MS-2 screening.

Conclusion

There was no effect of tube type on urine culture or MS-2 results when specimens were tested immediately upon arrival, but storage at room temperature for 24 h in BD tubes resulted in an increase in false positives by both MS-2 and culture.

Studies (1)

Purpose

This study investigated the effects of transporting and storing specimens in BD rather than conventional tubes on results of urine culture and MS-2 screening. Midstream urine from 312 obstetrics patients was aliquoted into BD urine culture tubes containing boric acid-glycerol-sodium formate and sterile containers and transported to the laboratory within 20 min or refrigerated until transport. After arrival, specimens were immediately cultured and analyzed using MS-2 screening and specimens in BD tubes were stored at room temperature and analyzed again 24 h later. Isolates were identified when colony counts exceeded 50,000 gram-positive cocci per mL or 100,000 gram-negative bacilli per mL. The time until bacteriauria was recorded for the MS-2 and the specimen was considered negative when no growth was within 6 h. 

Summary of Findings

Of the 312 specimens collected, 124 were positive using MS-2 screening in both the conventional and BD tubes and the average time to be determined positive was also comparable. Specimens that were negative by MS-2 had comparable colony counts for specimens in BD tubes and conventional tubes, but different when specimens were stored at room temperature for 24 h in BD tubes before culture there was an increase in positive cultures. Twenty-one of the 188 MS-2 negative specimens (14 in conventional tubes and 8 in BD tubes) grew >50,000 colonies/mL when cultured immediately and a similar number of specimens contained gram-positive cocci. Forty of the 188 MS-2 negative specimens grew >50,000 colonies/mL 24 h storage at room temperature in BD tubes before culture and 19 of these would have been considered to have significant growth. Importantly, specimens with yeast in conventional tubes also had yeast in BD tubes, but in smaller numbers.

  • Biospecimens

    FluidUrine
  • Preservation Types

    None (Fresh)Other Preservative
  • Diagnoses

    Pregnant
  • Platforms

    Cell count/volumeMicrobiological assay
    Cell count/volumeLight scattering
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    Biospecimen PreservationType of fixation/preservation
    Boric acid
    Glycerol
    Sodium formate
    StorageTime at room temperature
    0 h
    24 h
    Microbiological assay SpecificTechnology platform
    MS-2
    Urine culture