Assessment of a four hour delay for urine samples stored without preservatives at room temperature for urinalysis.

Veljkovic K, Rodríguez-Capote K, Bhayana V, Pickersgill R, Beattie J, Clark L, Kavsak PA

Publication: Clin Biochem (2012) • Vol 45 • Page 856-8PMID: 22548911View on PubMed

Purpose

This paper investigated the effects of storing specimens at room temperature for 4 h rather than 2 h on the results of chemical and microscopy analysis of urine.

Conclusion

Greater than 97% concordance in results was achieved between specimens analyzed within 2 h and 4 h for all analytes and complete agreement of results was obtained in 54% or greater in all specimens.

Studies (1)

Purpose

This study investigated the effects of storing specimens at room temperature for 4 h rather than 2 h on the results of chemical and microscopy analysis of urine. Levels of nitrites, glucose, leukocyte esterase, hemoglobin, proteins, ketones, specific gravity, pH, epithelial cells, erythrocytes, bacteria, and leukocytes were determined in 83 specimens after 2 h at room temperature and again after 4 h. Concordance was defined as results within one color block from the best-fit analysis or within one reporting range. Agreement was defined as both results on the best-fit line.

Summary of Findings

Concordance of glucose, leukocytes esterase, hemoglobin, proteins, and epithelial cell counts was 100% between specimens analyzed within 2 h and after 4 h. Further, 99% concordance for ketones and specific gravity, 98% for pH, and 97% for erythrocytes, bacteria, and leukocyte counts was observed for specimens analyzed after 2 and 4 h, respectively. Importantly, complete agreement of results was obtained in 54% of specimens for pH, 76% of specimens for specific gravity, 77% of specimens for leukocyte counts, 86% of specimens for proteins, 87% of specimens for epithelial cell counts, 88% of specimens for hemoglobin, 90% of specimens for bacterial counts, 91% of specimens for leukocyte esterase, 93% of specimens for ketones and erythrocyte counts, 99% of specimens for glucose, and 100% of specimens for nitrites.

  • Biospecimens

    FluidUrine
  • Preservation Types

    None (Fresh)
  • Diagnoses

    Not specified
  • Platforms

    CarbohydrateClinical chemistry/auto analyzer
    Cell count/volumeLight microscopy
    Electrolyte/MetalClinical chemistry/auto analyzer
    ProteinClinical chemistry/auto analyzer
    Small moleculeClinical chemistry/auto analyzer
  • Pre-analytical Factors

    ClassificationPre-Analytical FactorValue(s)
    StorageTime at room temperature
    2 h
    4 h