This paper investigated the optimization of urine analysis by MALDI-TOF MS and examined the effects of storage of urine and diagnosis of prostate cancer on the protein profile.
Optimization of analytical and pre-analytical conditions for MALDI-TOF-MS human urine protein profiles.
Calvano CD, Aresta A, Iacovone M, De Benedetto GE, Zambonin CG, Battaglia M, Ditonno P, Rutigliano M, Bettocchi C
Purpose
Conclusion
The largest number of peaks and highest peak intensities were obtained when urine was prepared using homemade hydrophilic–lipophilic balance (HLB) cartridges and was deposited by spin coating at a 1:1 ratio with the CHCA matrix. The protein profile was less affected by storage at -20˚C than 4˚C. Although the profile was altered after only two days of storage at 4˚C, effects of storage at -20˚C were first observed after 35 days.
Studies (2)
Purpose
This study investigated the effect of sample preparation method as well as matrix type, matrix to analyte ratio, and sample deposition on the MALDI-TOF MS spectra of a single urine specimen. The effect of prostate cancer on the protein profile was also investigated. Mid-stream early morning urine from a healthy patient was acidified to pH 3.0 with 0.5M phosphoric acid and centrifuged at 5000 x g for 10 min. Proteins were prepared with a ZipTip pipette tip, filtered through solid-phase extraction (SPE) using homemade columns containing R1 or R3 resin or OASIS hydrophilic–lipophilic balance (HLB) cartridges, or applied to one of 3 different Empore extraction disks (ANION SR, SDB-XC, or chelating). Prior to preparation using ZipTip, some aliquots were ultra-filtered using Vivaspin 500 10 kDa or 50 kDa MWCO. Optimization of MALDI-TOF analysis was conducted with proteins prepared using OASIS HLB cartridges. The effect of diagnosis was investigated using mid-stream early morning urine from 30 healthy individuals and 30 patients with prostate cancer processed according to the identified optimal protocol.
Summary of Findings
Urine that was not ultrafiltered before ZipTip preparation had few proteins of 1,000–12,000 Da, but the number of proteins detected increased slightly when the 50 kDa filter was used and larger increases were noted when the 10 kDa filter was used before ZipTip preparation. Importantly, the spectra obtained by the three methods were completely different. HLB resin cartridges resulted in more peaks in the spectra than columns packed with R1 or R3 resin or ANION SR, SDB-XC, or chelating extraction disks and also resulted in enrichment of the LMW peaks.
The highest peak intensities and largest number of peaks were obtained when CHCA was the matrix rather than SA or a mixture of CHCA and SA. A 1:1 urine to analyte matrix ratio was superior to 1:3 or 1:5 ratios. Sample deposition by spin coating rather than the dried droplet method resulted in the highest peak intensities and lowest standard deviation and coefficient of variance between spots.
Significant differences in urinary protein profiles were found among individuals with the same diagnosis making it hard to identify biomarkers. Further, no significant differences in the protein profile were identified between the pooled urine from 30 healthy individuals and the pooled urine from 30 patients with prostate cancer but some small differences were observed.
Biospecimens
FluidUrinePreservation Types
None (Fresh)Diagnoses
NeoplasticCarcinomaNormalPlatforms
ProteinMALDI-TOF MSPeptideMALDI-TOF MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Preaquisition Diagnosis/ patient condition HealthyProstate cancerAnalyte Extraction and Purification Analyte isolation method ANION SR Empore extraction diskChelating Empore extraction diskColumn containing R3 resinColumns containing R1 resinOASIS HLB cartridgeSDB-XC Empore extraction diskUltrafiltered using Vivaspin 500 10 kDa MWCO before ZipTipUltrafiltered using Vivaspin 500 50 kDa MWCO before ZipTipZipTip aloneMALDI-TOF MS Specific Resin CHCAMixture of CHCA and SASAMALDI-TOF MS Specific Template/input amount 1:1 ratio1:3 ratio1:5 ratio
Purpose
This study investigated the effects of storing urine at 4˚C or -20˚C on the MALDI-TOF MS spectra. Mid-stream early morning urine was collected from a healthy patient. After collection, urine was acidified to pH 3.0 with 0.5M phosphoric acid and centrifuged at 5000 x g for 10 min. Urine was aliquoted for storage at 4˚C for 2, 6, 11, 14, 21, or 24 days or at -20˚C for 2, 6, 11, 14, 35, 45, or 55 days. Urine was prepared using OASIS hydrophilic–lipophilic balance (HLB) cartridges and subjected to MALDI-TOF MS using the identified conditions. The abundance of the 70 most abundant ions between 1,000 and 12,000 Da in fresh urine with a signal to noise ratio of more than three was used to generate a peak list. The intensities of these peaks were compared between stored specimens and fresh specimens.
Summary of Findings
Variations in the protein profile were apparent after only two days of storage at 4˚C with loss of some high molecular weight peaks and a corresponding increase in low molecular weight proteins. Principle component analysis revealed that 95.62% of the variance in the spectra could be explained by the first two components: the difference between fresh specimens and those stored at 4˚C and two peaks that vanished after 2 and 14 days of storage.
Storage of urine at -20˚C led to better preservation of the spectra and most of the observed effects were seen on peaks lower than 5,300 Da. Principle component analysis revealed that 65.03% of the variance was attributable to differences in the intensities of peaks less than 5,300 Da and 16.47% was due to differences between specimens analyzed before 35 days and those analyzed after 45 or 55 days.
Biospecimens
FluidUrinePreservation Types
FrozenNone (Fresh)Diagnoses
NormalPlatforms
PeptideMALDI-TOF MSProteinMALDI-TOF MSPre-analytical Factors
Classification Pre-Analytical Factor Value(s) Biospecimen Preservation Type of fixation/preservation FrozenNone (fresh)RefrigerationStorage Storage duration 11 days14 days2 days21 days24 days35 days45 days55 days6 days